Connect the equipment.
1
Connect the signal generator to the input of the power splitter.
a
Connect the power sensor to one output of the power splitter, and
b
connect channel 1 of the oscilloscope to the other power splitter
output (put the 50 Ω feedthrough at the input of the oscilloscope).
Set up the oscilloscope.
2
, then press the Default Setup softkey.
Press
a
Setup
Set the time base to 500 ns/div.
b
Set channel 1 to 100 mV/div.
c
Press
, then press the Average softkey.
d
Display
Toggle the # Average softkey to select 8 averages.
e
Set the signal generator for 1 MHz at about 5.6 dBm.
3
Notice that the signal on the display is about 5 cycles and six divisions of
amplitude.
Verifying Oscilloscope Performance
Service
4–13