If you need to compare several single-shot events,
6
press
.
Autostore
Like the
key,
Run
the trigger conditions are met, the oscilloscope triggers. Pressing
again rearms the trigger circuit, but this time the display
Autostore
is not erased. All the data points are retained on the display in half
bright with each trigger allowing you to easily compare a series of
single-shot events.
After you have acquired a single-shot event, pressing a front-panel key,
softkey, or changing a knob can erase the event from the display. If you
press
, the oscilloscope will recover the event and restore the
Stop
oscilloscope settings.
To clear the display, press
To exit the Autostore mode, press either
or
. Notice that RUN replaces STORE in the status line,
Autostore
indicating that the oscilloscope has exited the Autostore mode.
Operating hint
The single-shot bandwidth is 2 MHz for single-channel operation, and 1 MHz for
two-channel operation. There are twice as many sample points per waveform
on the one-channel acquisition than on the two-channel acquisition. On the
HP 54600B and HP 54603B, channels 1 and 2 are captured simultaneously. On
the HP 54601B and HP 54602B channels 1 and 2 are captured simultaneously,
then on the next trigger channels 3 and 4 are captured simultaneously.
also arms the trigger circuit. When
Autostore
.
Erase
Run
Operating Your Oscilloscope
To capture a single event
2–9