To Verify Bandwidth (Alternate Method) - HP 54600 Series User's And Service Manual

Hide thumbs Also See for 54600 Series:
Table of Contents

Advertisement

Service
Verifying Oscilloscope Performance

To verify bandwidth (alternate method)

In this test you verify the bandwidth of the oscilloscope by using a power
meter and power sensor to set the output of a signal generator at 1 MHz and
the upper bandwidth limit. You use the peak-to-peak voltage at 1 MHz and
the upper bandwidth limit to calculate the bandwidth response of the
oscilloscope.
Test limits
HP 54600B and HP 54601B, all channels (−3 dB)
HP 54602B
HP 54603B, all channels (−3 dB)
Table 4-7
Equipment Required
Equipment
Signal generator
Power meter and
Power Sensor
Power splitter
Cable
Adapter
Feedthrough
*
4–12
dc to 100 MHz
ac coupled 10 Hz to 100 MHz.
Channels 1 & 2 (–3 dB) *
dc to 150 MHz
ac coupled 10 Hz to 150 MHz.
Channels 3 & 4 (–3 dB)
dc to 250 MHz.
dc to 60 MHz
ac coupled 10 Hz to 60 MHz.
Critical specifications
1 to 100 MHz at 200 mV
(250 MHz for HP 54602B)
1 to 100 MHz ±3% accuracy
(250 MHz for HP 54602B)
Outputs differ by <0.15 dB
Type N (m), 24 inch
Type N (m) to BNC (m)
50 Ω, BNC (m) and (f)
1mv/div to 5 mv/div dc = 100 MHz.
Recommended Model/Part
HP 8656B opt 001
HP 436A and HP 8482A
HP 11667B
HP 11500B
HP 1251-0082
HP 10100C