HP 54600B User, Installation And Service Manual page 104

Hide thumbs Also See for 54600B:
Table of Contents

Advertisement

Change the time base to 5 ns/div.
8
Wait a few seconds for the measurement to settle (averaging is complete),
then note the Vp-p reading from the bottom of the display.
Vp-p = ______ mV.
Calculate the response using the following formula.
9
20 log
Correct the result from step 9 with any difference in the power meter
10
reading from step 7. Make sure you observe all number signs.
Result from step 9 = −2.3 dB
For example
Power meter reading from step 7 = −0.2 dB (REL)
True response = (−2.3) − (−0.2) = −2.1 dB
The true response should be ≤±3 dB.
If the result is not ≤±3 dB, see "Troubleshooting the Oscilloscope," on
page 4–40.
Repeat steps 1 to 10 for channel 2 (channels 2 to 4 on the HP 54601B
11
and HP 54602B).
When you measure the bandwidth on channels 3 and 4, use the 0.1 V/Div
range.
step 8 result
10
step 4 result
Verifying Oscilloscope Performance
Service
4–15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 54601bHp 54602bHp 54603b

Table of Contents