Performance Characteristics
Option 005 Trigger System
Option 005 Trigger System
Internal trigger
Sensitivity Performance remains unchanged
Coupling Performance remains unchanged
Modes Performance remains unchanged
Holdoff Performance remains unchanged
TV triggering Available on channels 1 and 2 only
TV line and field 0.5 division of composite sync for stable display
External trigger
Performance remains unchanged
Vertical output
Connector Rear panel BNC (f)
Source Impedance 50Ω (nominal)
Signal source selected by internal trigger source
Amplitude approximately 90mVp-p into 50Ω for a full scale display at full
bandwidth of the oscilloscope
TV Trigger output
Connector Rear panel BNC (f)
Amplitude TTL
Pulse width a function of TV trigger mode, Minimum approximately 5us in
line modes to the width of a field in field modes
Delay from Vertical Output approximately 400ns.