HP 54600B User, Installation And Service Manual page 100

Hide thumbs Also See for 54600B:
Table of Contents

Advertisement

1
Using the 50-Ω feedthrough and the
generator to channel 1 of the oscilloscope.
Set the frequency of the signal generator to 250 kHz and the
2
amplitude to about 800 mV.
Press
3
Adjust the output of the signal generator for exactly 8 divisions of
4
vertical deflection.
Change the frequency of the signal generator to the value shown
5
below for your instrument.
Table 4-6
Signal Generator Frequency Setting
Selected
Channel
Channel 1
Channel 2
Channel 3
Channel 4
*1mv/div to 5 mv/div dc = 100 MHz.
Change the sweep speed of the oscilloscope to 5 ns/div and observe
6
the display.
The vertical amplitude of the signal on the display should be equal to or
greater than 5.66 divisions (−3 dB point). If the result is not ≤−3 dB, see
"Troubleshooting the Oscilloscope," on page 4–40.
Repeat steps 1 through 6 for channel 2 (channels 2 to 4 on the
7
HP 54601B and HP 54602B).
.
Autoscale
HP 54600B
HP 54601B
100 MHz
100 MHz
100 MHz
100 MHz
100 MHz
100 MHz
Verifying Oscilloscope Performance
cable, connect the signal
BNC
HP 54602B *
HP 54603B
150 MHz
60 MHz
150 MHz
60 MHz
250 MHz
250 MHz
Service
4–11