Service
Verifying Oscilloscope Performance
Table 4-11
Signal Generator Frequency Setting
Selected
Channel
Channel 1
Channel 2
Channel 3
Channel 4
*1mv/div to 5 mv/div = 100 MHz.
Press
b
Decrease the output of the signal generator until there is 1 vertical
c
division of the signal displayed.
The trigger should be stable. If the triggering is not stable, try
adjusting the trigger level. If adjusting the trigger level makes the
triggering stable, the test still passes. If adjusting the trigger does not
help, see "Troubleshooting the Oscilloscope," on page 4–40.
Repeat steps 1 through 4 for channel 2 on the HP 54600B and
5
HP 54603B (channels 2 to 4 on the HP 54601B and HP 54602B).
4–20
HP 54600B
HP 54601B
100 MHz
100 MHz
100 MHz
100 MHz
—
100 MHz
—
100 MHz
.
Autoscale
HP 54602B *
HP 54600B
150 MHz
60 MHz
150 MHz
60 MHz
250 MHz
—
250 MHz
—