Vega VEGAPULS 63 Operating Instructions Manual page 20

Radar sensor for continuous level measurement of aggressive liquids or with hygienic requirements 20 ma/hart - four-wire
Hide thumbs Also See for VEGAPULS 63:
Table of Contents

Advertisement

4 Mounting
20
100%
0%
Fig. 12: VEGAPULS 63 in a bypass tube. The polarisation marking on the
process fitting must be in one plane with the tube holes or the tube connection
openings.
1
Marking of the polarisation direction
When the sensor is mounted on a bypass tube, the distance
from VEGAPULS 63 to the upper tube connection should be
approx. 300 mm (11.81 in) or more. In case of extremely rough
tube inner walls, you should use an inserted tube (tube in tube)
or a radar sensor with tube antenna.
Information:
With VEGAPULS 63 in flange version, the polarisation plane is
always in the center between two flange holes.
VEGAPULS 63 • 4 ... 20 mA/HART - four-wire