Vega VEGAPULS 61 Operating Instructions Manual page 61

Radar sensor for continuous level measurement of liquids
Hide thumbs Also See for VEGAPULS 61:
Table of Contents

Advertisement

Code
Text message
F260
Error in the calibration
F261
Error in the configuration
F264
Installation/Setup error
F265
Measurement function dis-
turbed
Function check
Out of specification
VEGAPULS 61 • Foundation Fieldbus
Cause
– Error in the calibration
carried out in the factory
– Error in the EEPROM
– Error during setup
– False signal suppression
faulty
– Error when carrying out
a reset
– Adjustment not within
the vessel height/meas-
uring range
– Max. measuring range
of the instrument not
sufficient
– Sensor no longer carries
out a measurement
– Operating voltage too
low
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text message
C700
Simulation active
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Code
Text message
S600
Impermissible
electronics tem-
perature
9 Diagnosis, asset management and service
Rectification
– Exchanging the elec-
tronics
– Send instrument for
repair
– Repeat setup
– Repeat reset
– Check or correct instal-
lation and/or parameter
adjustment
– Use an instrument with
bigger measuring range
– Check operating voltage
– Carry out a reset
– Disconnect operating
voltage briefly
Cause
Rectification
– A simulation is
– Finish simula-
active
tion
– Wait for the
automatic end
after 60 mins.
Cause
Rectification
– Temperature of
– Check ambient
the electronics
temperature
in the non-spec-
– Isolate electron-
ified range
ics
– Use instrument
with higher
temperature
range
DevSpec
Diagnosis Bits
Bit 8
Bit 9
Bit 10
Bit 11
DevSpec Diag-
nosis Bits
Bit 19
DevSpec Diag-
nosis Bits
Bit 18
61