Vega VEGAPULS 61 Quick Setup Manual

Radar sensor for continuous level measurement of liquids, profibus pa
Hide thumbs Also See for VEGAPULS 61:

Advertisement

Quick setup guide

Radar sensor for continuous level
measurement of liquids
VEGAPULS 61
Profibus PA
Document ID: 51870

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VEGAPULS 61 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Vega VEGAPULS 61

  • Page 1: Quick Setup Guide

    Quick setup guide Radar sensor for continuous level measurement of liquids VEGAPULS 61 Profibus PA Document ID: 51870...
  • Page 2: Table Of Contents

    Operating Instructions Manual as well as the Safety Manual that comes with instruments with SIL qualification. These manuals are available in the download area of "www.vega.com". Operating instructions VEGAPULS 61 - Profibus PA: Document- ID 28444 Editing status of the quick setup guide: 2018-02-16...
  • Page 3: For Your Safety

    During work on and with the device the required personal protective equipment must always be worn. Appropriate use VEGAPULS 61 is a sensor for continuous level measurement. You can find detailed information about the area of application in chapter "Product description".
  • Page 4: Safety Label On The Instrument

    For operation inside of closed vessels, points a to f in annex E of EN 302372 must be fulfilled. FCC/IC conformity (only for USA/Canada) VEGAPULS sensors with all antenna versions are FCC/IC approved. VEGAPULS 61 • Profibus PA...
  • Page 5: 1.10 Environmental Instructions

    Modifications not expressly approved by VEGA will lead to expiry of the operating licence according to FCC/IC. VEGAPULS 61 is in conformity with part 15 of the FCC directives and fulfills the RSS-210 regulations. Note the corresponding regulations for operation: •...
  • Page 6: Product Description

    Alternatively, you can access the data via your smartphone: • Download the VEGA Tools app from the "Apple App Store" or the "Google Play Store" • Scan the Data Matrix code on the type label of the instrument or •...
  • Page 7: Mounting

    > 10 mm into the vessel > 200 mm (7.87") Fig. 1: Distance of the antenna to the vessel wall/vessel ceiling 2. Note min. socket diameter depending on the socket length For further information see chapter "Mounting". VEGAPULS 61 • Profibus PA...
  • Page 8: Connecting To Power Supply

    11. Tighten the compression nut of the cable entry gland. The seal ring must completely encircle the cable 12. Screw the housing lid back on The electrical connection is finished. VEGAPULS 61 • Profibus PA...
  • Page 9: Wiring Plan, Single Chamber Housing

    Wiring plan, double chamber housing The following illustrations apply to the non-Ex as well as to the Ex-ia version. Wiring plan I 2 C Fig. 4: Wiring plan - double chamber housing Voltage supply, signal output VEGAPULS 61 • Profibus PA...
  • Page 10: Set Up With The Display And Adjustment Module Plicscom

    Fig. 5: Insert display and adjustment module Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher lid with an inspection glass is required. VEGAPULS 61 • Profibus PA...
  • Page 11: Setup Steps

    The actual level is then calculated on the basis of these entered values. At the same time, the operating range of the sensor is limited from maximum range to the requested range. VEGAPULS 61 • Profibus PA...
  • Page 12 1. Move from the measured value display to the main menu by pushing [OK]. 2. Select the menu item "Service" with [->] and confirm with [OK]. Now the menu item "False signal suppression" is displayed. VEGAPULS 61 • Profibus PA...
  • Page 13: Menu Schematic

    1.000 m(d) 8.000 m(d) 2.000 m(d) Vessel form Linearisation curve Channel Damping ▼ Linear PV lin. value not known Sensor-TAG Sensor Display Basic adjustment ▶ Display Diagnostics Service Info Displayed value Backlight ▼ PA-Out Switched off VEGAPULS 61 • Profibus PA...
  • Page 14 Language Copy sensor data ▼ Copy sensor data? Enable? Deutsch Info Basic adjustment Display Diagnostics Service ▶ Info Instrument type Date of manufacture Last change using PC Sensor characteristics Software version Display now? Serial number VEGAPULS 61 • Profibus PA...
  • Page 15: Supplement

    Ʋ Ex ia instrument 12 … 24 V DC Ʋ Ex-d-ia instrument Lighting not possible Power supply by DP/PA segment coupler Depending on the version M12 x 1, according to DIN 43650, Harting, 7/8" FF. VEGAPULS 61 • Profibus PA...
  • Page 16 6 Supplement Max. number of sensors non-Ex/Ex 32/10 VEGAPULS 61 • Profibus PA...
  • Page 17 Notes VEGAPULS 61 • Profibus PA...
  • Page 18 Notes VEGAPULS 61 • Profibus PA...
  • Page 19 Notes VEGAPULS 61 • Profibus PA...
  • Page 20: Information

    Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2018 VEGA Grieshaber KG Phone +49 7836 50-0 Am Hohenstein 113...