JUMO LOGOLINE 500d Operating Manual page 66

Pen recorder with text printing and led dot matrix display
Table of Contents

Advertisement

Display
Edit
(Continued:
INPUT)
(THERMO-
COUPLE)
(only with
EXT.
CONST.)
1. The cold junction temperature of the internal cold junction is measured on channel 1. It is also valid for channels 2 and 3.
If one or more thermocouple inputs are configured with an internal cold junction, a thermocouple with internal cold junction
measurement or external constant cold junction temperature has to be configured for channel 1.
8 Programming
Selection/input
1 CHARACT.:L
4
L
(–200 to +900°C)
J
(–210 to +1200°C)
U
(–200 to +600°C)
T
(–270 to +400°C)
K
(–270 to +1372°C)
E
(–270 to +1000°C)
N
(–270 to +1300°C)
S
(–50 to +1768°C)
R
(–50 to +1768°C)
B
X1
(customized lin. 1)
X2
(customized lin. 2)
1 UNIT: °C
41
°C, °F
1 RANGE BEG: –200.0 °C
42
Input start of
measurement range
1 RANGE END: +900.0 °C
43
Input end of
measurement range
1 TEMP . COMP:
44
INTERN Pt 100
INTERN Pt 100
EXTERN Pt 100
EXT. CONST.
1 COMP . TEMP .:
45
+0020. °C
Input cold junction
temperature
(value range:
–20 to +100°C)
with keys Enter
h
R
,
(0 — 1820°C)
h
R
,
h
R
,
,
l
r
,
,
S
h
R
,
,
l
r
,
,
S
h
R
,
h
R
,
,
l
r
,
,
S
66
Continue with
E
customized
X1 or X2:
10 (FILTER)
otherwise:
41
E
42
E
43
Check
range start
according to
characteristic
E
44
Check
range end
according to
characteristic
end – start
100°C
S,R,B
500°C
E
INTERN Pt 100
EXTERN Pt 100:
10 (FILTER)
EXT. CONST.:
45
E
10 (FILTER)
Check
value range
1
: