Service; Converter Check List - Siemens MAG 1100 Handbook

Electromagnetic flowmeters
Table of Contents

Advertisement

SITRANS F M MAGFLO

9. Service

9.1

Converter check list

YES
Power on converter,
display light on
NO
Check cables/connections
YES
Check connection board
Check pins in converter
multiplug - OK
NO
Correct fault
74
9. Service
Often problems with unstable/wrong measurements occur due to insufficient/wrong earthing or
potential equalization. Please check this connection. If OK, the SITRANS F M MAGFLO
verter can be checked as described under 9.1 and sensor under 9.3.
When checking SITRANS F M MAGFLO
check the signal converter is to replace it with another MAG 5000/6000 converter with a similar
power supply.
A replacement can easily be done as all settings are stored in and downloaded from the
SENSORPROM
unit - no extra settings need to be made.
If no spare converter is available - then check converter according to check table.
YES
Output
Display defect
readings OK
Change display
NO
Converter
defective
DKFD.PS.027.W3.02
installations for malfunction the easiest method to
YES
Error triangles flashing
NO
YES
Output and display
readings OK ?
NO
Check cables/connections
YES
Check connection board
Check pins in converter
multiplug - OK
NO
Correct fault
con-
Check error table
Converter OK -
Check settings/application
Check installation/sensor/
earthing connection etc.