Installation Of Signal Converter; Compact Installation Mag 5000 And Mag 6000 - Siemens MAG 1100 Handbook

Electromagnetic flowmeters
Table of Contents

Advertisement

SITRANS F M MAGFLO
6. Installation of signal
converter
6.1
Compact installation
MAG 5000 and MAG 6000
Compact polyamide
Note
System will not register flow
if black plugs are not
connected to connection
board

6. Installation of signal converter

DKFD.PS.027.W3.02
Step 1
Remove and discard the terminal box lid of the
sensor.
Fit the PG 13.5 cable glands for the supply and
output cables.
Step 2
Remove the two black plug assemblies for coil
and electrode cables in the terminal box and
connect them to their corresponding terminal
numbers on the connection board.
Step 3
Connect an earth wire between PE on connec-
tion board and bottom of connection box.
Connect the 2 pin connector and 3 pin connector
as shown.
Note
In earlier version the 3 pin connector was a 5 pin
connector.
Step 4
Mount the connection plate in the terminal box.
®
The SENSORPROM
unit connections will be
established automatically when the connection
plate is mounted in the terminal box.
Note
Check that your connection board lines up with
®
the SENSORPROM
unit, if not, move the
®
SENSORPROM
unit to the other side of the
terminal box.
Step 5
Fit the supply and output cables respectively
and tighten the cable glands to obtain optimum
sealing.
Please refer to the wiring diagram in chapter 7
for the electrical connections.
Mount the signal converter on the terminal box.
39