Specifications

Standard Features

Feature
Model name
Measuring method
Measurement range
RMSE
Light source
Light source life
Sensors
Power source
Protection type and level
Minimum number of
measurements when fully charged
Operating temperature range
Operating relative humidity range
Storage temperature range
Storage relative humidity range
Dimensions
Weight, including battery
AC adapter input for North America 120V, 50/60 Hz, 10 W
AC adapter input for
International use
AC adapter output
* This accuracy is based on the aggregate results from both clinical
study groups (n=613). Refer to Appendix A for a detailed description of
results by clinical site, measurement location, and patient
demographics.
Page 2 - 10
JM-103
Determines the yellowness of the
subcutaneous tissue by using two
optical paths to measure the optical
density difference at two wave-
lengths
0.0 mg/dL to 20 mg/dL or 0 µmol/L
to 340 µmol/L
± 1.5 mg/dL or ± 25.5 µmol/L
Pulse xenon arc lamp
150000 measurements
Silicon photodiodes
Special Ni-MH battery
Internally-powered instrument, BF-
type
400 single measurements
10°C (50°F) to 40°C (104°F)
30% to 95%
-10°C (14°F) to 50°C (122°F)
30% to 95%
48 mm (1.9") wide x 15.5 cm (6.1")
high x 33 mm (1.3") deep
150 g (5.3 oz)
200V-240V, 50/60 Hz, 12.5VA
9V, 500 mA