Cleaning
WARNING:
Follow the product manufacturer's instructions. Failure to do so
could result in personal injury or equipment damage.
SHOCK HAZARD:
Before cleaning, maintenance, or parts replacement, unplug the
charger unit from its power source. Failure to do so could result in
personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture that would allow for
liquid pooling. Personal injury or equipment damage could occur.
CAUTION:
Do not use harsh cleansers/detergents, such as scouring pads
and heavy duty grease removers, or solvents, such as toluene,
xylene, and acetone. Equipment damage could occur.
If there is no visible soilage with possible body fluids, we recommend
that you clean the unit with a mild detergent and warm water. If
disinfection is desired, you may use a combination cleanser/disinfectant
as explained in "Disinfecting" on page 6-2.
Steam Cleaning
Do not use any steam cleaning device on the unit. Do not autoclave the
unit. Excessive moisture can damage mechanisms in this unit.
Cleaning Hard to Clean Spots
To remove difficult spots or stains, we recommend that you use standard
household cleansers and a soft bristle brush. To loosen heavy, dried-on
soil, you may first need to saturate the spot.
Jaundice Meter (Model JM-103) User Manual (usr070rb)
Cleaning, Maintenance,
and Replacement Parts
Section 6
Page 6 - 1