Operational Checkout of the Jaundice Meter

Do not touch the checker's surface with your fingers. If the checker gets
dirty, wipe it with a soft cloth dampened with water, and then wipe it
with a dry cloth.
Using the checker supplied with the charger unit, check the instrument
to ascertain that the meter light output is within range. Check the
instrument once or more each day it is used. To inspect the Jaundice
Meter, perform the following:
1.
Hold the Reset button down, and set
the Power switch to the On position.
NOTE:
If the Reset button is held down for longer
than 15 seconds, the unit of measurement
switches.
2.
After CHE appears on the display
window, immediately release the
Reset button. If the Reset button is
held down for longer than 15 seconds,
switch the unit of measurement back to
its previous setting (see "Selecting the
Unit of Measurement" on page 4-3).
3.
Visually confirm that CHE appears in
the display and that the Ready lamp
illuminates.
4.
Open the cover of the checker. Use
only the checker supplied with the
Jaundice Meter (JM-103).