Service Calls

When you call Hill-Rom about your unit, be prepared to give the serial
number from the product identification label. When you give the serial
number, the Hill-Rom representative can identify your unit and give you
the information you need more quickly.

Error Messages

For warnings that may appear on the display window, refer to the table
below.
Warning
Er1
The measured value is
abnormal. In the case of an
averaging measurement, the
measurement fluctuation is
excessively large.
Er2
A measurement error may
through
have occurred during an aver-
Er6
aging measurement, or the
hardware is not functioning
properly.
Jaundice Meter (Model JM-103) User Manual (usr070rb)
Error Messages
Cause
Section 7
Troubleshooting
Solution
Place the measuring probe
perpendicular to the infant's
forehead or sternum, and take
the measurement again. If
Er1 still appears, contact the
nearest authorized service
facility.
Set the Power switch to the
Off position, and then return
it to the On position. If the
warning still appears, contact
the nearest authorized service
facility.