In The Event Of Problems - Beurer BGL40 Instructions For Use Manual

Blood pressure and blood glucose monitor two in one
Hide thumbs Also See for BGL40:
Table of Contents

Advertisement

10 I
N THE EVENT OF PROBLEMS
Messages on the display concerning batteries and blood glucose testing
No.
Cause
E-b
Low batteries
E-t
Temperature of
environment, monitor or
test strip outside approved
range
E-U
Used or contaminated test
strip was inserted
Messages on the display concerning blood pressure testing
No.
Cause
Err 01
Cuff leaks or pressure too
low
Err 02
Systolic pressure cannot
be measured
Err 03
Cuff pressure is too high
(greater than 300 mmHg)
Err 04
Diastolic pressure cannot
be measured
Err 07
Systolic pressure out of
range
Beurer BGL40 mg/dL
Remedy
Replace all batteries.
Repeat test with new test strip once
environment, monitor and test strip are
at room temperature (+20 to +25°C).
• Insert an unused and unexpired test
strip
• Repeat blood glucose test
Remedy
Position the cuff correctly (see "5.1
Positioning the cuff", page 21) and
repeat the test
Repeat test and do not move or talk
during the test.
Do not crimp the cuff.
Repeat test and do not move or talk
during the test.
Repeat test and do not move or talk
during the test.
41