  • Page 1 BGL40 BLOOD PRESSURE AND BLOOD GLUCOSE MONITOR TWO IN ONE Step by step Beurer GmbH & Co. KG. Söflinger Straße 218. 89077 Ulm. Germany. www.beurer-medical.de...
  • Page 3: Table Of Contents

    Contents of BGL40 mg/dL Blood Pressure and Blood Glucose Monitor Getting to know your instrument..........2 1.1 Contents and purchasing supplies ............2 1.2 Monitor functions................. 4 1.3 Health information ................5 1.4 Explanation of symbols ............... 6 Warning and safety information ..........8 Description of monitor and accessories........
  • Page 4: Getting To Know Your Instrument

    1.1 Contents and purchasing supplies Check that the external packaging of the Beurer BGL40 mg/dL monitor is undamaged and that all components are included.
  • Page 5 You can obtain test strips, control solution, lancets and lancing device without a prescription. Article 50 test strips REF 457.00 control solution MEDIUM REF 457.02 control solution HIGH REF 457.03 control solution LOW REF 457.04 100 lancets REF 457.01 Beurer BGL40 mg/dL...
  • Page 6: Monitor Functions

    • display of the ambient temperature and warning of unsuitable temperatures • 'change battery' display when batteries are low Note • Do not use the monitor to diagnose diabetes but for regular blood glucose monitoring only. • Consult your doctor to set insulin dosages. Beurer BGL40 mg/dL...
  • Page 7: Health Information

    1 hour after meals Below 160 mg/dL 2 hours after meals Below 120 mg/dL Between 2 and 4 a.m. Over 70 mg/dL Source: Krall, L.P., and Beaser, R.S :Joslin Diabetes Manual. Philadelphia: Lea and Febiger (1989),1 8 Beurer BGL40 mg/dL...
  • Page 8: Explanation Of Symbols

    CIS countries Green Dot: German Maximum storage time dual waste disposal in months after opening system Contents sufficient for Batch <n> tests Sterilised by irradiation Order number /Art.-Nr. (lancets) Attention! Biohazard, danger of Read accompanying infection documentation Beurer BGL40 mg/dL...
  • Page 9 When measuring blood pressure do not talk. Blood glucose measurement unit The symbols in the Instructions for Use denote the following: WARNING Warning of risk of injury or health hazards CAUTION Safety information about possible damage to monitor and accessories Note Important information Beurer BGL40 mg/dL...
  • Page 10: Warning And Safety Information

    • The readings that you take may only be used for information purposes – they are not a substitute for a medical examination! Discuss your results with the doctor. Never use them to make medical decisions independently (e.g. regarding medication and dosage). Beurer BGL40 mg/dL...
  • Page 11 • Lipaemia effects: Elevated blood triglycerides up to 2000 mg/dL have virtually no influence on the results. However, triglycerides above this level may influence the blood glucose test results. • Use only fresh whole capillary blood. Do not use serum or plasma. Beurer BGL40 mg/dL...
  • Page 12 • Do not use the test strips at altitudes above 3275 m. Note The Beurer BGL40 mg/dL monitor is suitable for measuring whole capillary blood. Storage and care WARNING •...
  • Page 13 • Never open the monitor. If these instructions are not heeded, the warranty becomes null and void. • Never attempt to repair the monitor yourself. We can no longer guarantee perfect functioning if you do. • If the monitor requires repair, please send it to our Customer Service. Beurer BGL40 mg/dL...
  • Page 14 Cd = battery contains cadmium, Hg = battery contains mercury. • The monitor should be disposed of according to Regulation 2002/96/EC-WEEE (Waste Electrical Electronic Equipment). In case of queries, please contact the municipal authorities responsible for waste disposal in your area. Beurer BGL40 mg/dL...
  • Page 15: Description Of Monitor And Accessories

    1: Wrist cuff 2: LCD display 3: ON/OFF button (Do not press button until the cuff is in position.) 4: Save button 'M' 5: Slot for test strips 6: Code button ‘C’ 7: 'SET' button 8: Battery compartment Beurer BGL40 mg/dL...
  • Page 16 7: High value (high) 16: Blood pressure unit mmHg 8: Low value (low) 17: Blood pressure unit kPa 9: Heart beat symbol 18: Diastolic blood pressure (DIA) 10: Pulse display, 19: Memory symbol average blood glucose 20: Display DIA Beurer BGL40 mg/dL...
  • Page 17 29: Temperature symbol 24: Replace battery symbol 25: DAY AVG for average reading 7,14,21,28,60,90 days Note The monitor is supplied with the following basic settings: • Blood pressure unit: mmHg • Blood glucose unit: mg/dL • Temperature unit: °C Beurer BGL40 mg/dL...
  • Page 18: Test Strips

    Read carefully the following information on the handling and storage of your test strips. All instructions must be followed to ensure that the test strips will give accurate test results. WARNING A test strip must be used only once and on only one patient. Beurer BGL40 mg/dL...
  • Page 19 Keep test strips away from direct sunlight or heat. Do not store in a refrigerator. • The relative humidity must be between 10% and 85%. • Store test strips in their original container – never place them in a different container. Beurer BGL40 mg/dL...
  • Page 20: Operation And Basic Settings

    Replace the battery cover carefully If the 'change battery' symbol appears, the battery is almost flat. Replace all batteries as soon as possible. The display shows E-b if the batteries are so flat that readings cannot be taken. Beurer BGL40 mg/dL...
  • Page 21: Making Basic Settings

    Setting blood pressure unit Press the 'M' button to switch between mmHg and kPa. Press the 'SET' button. The temperature unit starts blinking. Setting temperature unit Press the 'M' button to switch between °C and °F. Press the 'SET' button. Beurer BGL40 mg/dL...
  • Page 22: Measuring Blood Pressure

    • Avoid coffee, tea, alcohol and cigarettes for at least 30 minutes before testing your blood pressure. • Wait about 30 minutes before testing after exercise and swimming. • Blood pressure will naturally vary slightly between your two hands. Always measure your blood pressure on the same wrist. Beurer BGL40 mg/dL...
  • Page 23: Positioning The Cuff

    1 cm below the ball of your thumb. The cuff has to be fitted tightly around the wrist but should not constrict it. CAUTION Always position the cuff before you start the test. Otherwise the monitor may be damaged. Beurer BGL40 mg/dL...
  • Page 24: Correct Posture

    “5.4 Evaluating blood pressure results” page 23. If an error message appears, read section “10 In the event of problems” page 41. Press the ON/OFF button . If you forget to switch off the monitor, it switches off automatically after approx. three minutes. Beurer BGL40 mg/dL...
  • Page 25: Evaluating Blood Pressure Results

    105 mmHg • Diastolic: below 60 mmHg Regular self-monitoring = normal • Systolic: 105-140 mmHg • Diastolic: 60-90 mmHg Have it checked by your high doctor • Systolic: over 140 mmHg • Diastolic: over 90 mmHg Beurer BGL40 mg/dL...
  • Page 26: Ensuring Blood Glucose Measurability

    Press and hold the button to make the process faster. After code 75 the monitor continues to count from code 51. Save the new code number with the 'M' button. The monitor switches off automatically. Beurer BGL40 mg/dL...
  • Page 27: Checking Test Strips With Control Solution

    CAUTION Never use control solution from other manufacturers. It will be impossible to check that your Beurer blood glucose monitor operates correctly with other control solutions. When is a test with control solution recommended? •...
  • Page 28 (see "3.3 Test strips" page 16). The monitor switches on automatically and shows consecutively the initial display, the ambient temperature, and the code number. The monitor is ready for a test as soon as the flashing blood drop symbol is displayed. Beurer BGL40 mg/dL...
  • Page 29 95% of all tests. WARNING The results range printed on the test strip container is only applicable to the control solution. It is not a recommended value for your blood glucose level. Beurer BGL40 mg/dL...
  • Page 30 A problem with the monitor Contact Customer Service department. WARNING If you receive repeated results outside the specified range with the control solution, do not continue using the system to measure your blood glucose level. Contact our Customer Service department. Beurer BGL40 mg/dL...
  • Page 31: Measuring Blood Glucose

    • Blood samples from the fingerprints and samples from other parts of the body (AST) may have significantly different results. Always consult your doctor before starting testing at other parts of the body. Beurer BGL40 mg/dL...
  • Page 32 Under compression the blood will be diluted with tissue fluid and the test result may be inaccurate. • Note that poor circulation at the puncture site, which may be the result of cold or disease, may lead to inaccurate results. Beurer BGL40 mg/dL...
  • Page 33 • 1 to 2: soft or thin skin • 3: normal skin • 4 to 5: thick or calloused skin Rotate the cap in the appropriate direction until the arrow indicates the desired puncture depth. Beurer BGL40 mg/dL...
  • Page 34 Maintain the pressure until the drop has a volume least 0.5 microlitre volume (approx. 1.2 mm, original size: ). Carefully remove the lancing device from the skin. Beurer BGL40 mg/dL...
  • Page 35 Carefully dispose of all blood samples and materials that have come into contact with them. This will prevent injury and infection of other people. Screw the cap (grey) on again. Beurer BGL40 mg/dL...
  • Page 36: Measuring Blood Glucose

    Remove the test strip from the monitor and dispose of it as directed to prevent infection of other persons. Note • Do not apply more blood if the monitor does not start the measurement. Remove the test strip and stop the test. Use a new test strip. Beurer BGL40 mg/dL...
  • Page 37: Evaluating Blood Glucose Results

    600 mg/dL. Note If you suspect that the results are incorrect, first repeat the test and if necessary carry out a function test with control solution. If the questionable results continue, ask your doctor. Beurer BGL40 mg/dL...
  • Page 38 Ask your doctor. ketones 240 mg/dL and above Hi = severe hyperglycaemia Repeat test with a new test (very high blood sugar). strip. If it is the same as Over 600 mg/dL before, seek medical treatment immediately. Beurer BGL40 mg/dL...
  • Page 39: Memory

    • If values are already stored and you reset the date, the averages will be calculated by the new date. • "---" shows that the memory is empty. OFF is displayed briefly and the monitor automatically switches off. Beurer BGL40 mg/dL...
  • Page 40: Displaying Single Values

    You can display the average values of your blood glucose readings for the last 7, 14, 21, 28, 60 and 90 days. At the same time the monitor shows how frequently you have conducted tests over this period. Beurer BGL40 mg/dL...
  • Page 41: Deleting Stored Values

    The monitor must be switched off. Press the 'SET' button eight times slowly. Press ‘M’. dEL and the memory symbol blink on the display. Press 'M' again. CLr ALL is displayed briefly, then OFF and the monitor automatically switches off. Beurer BGL40 mg/dL...
  • Page 42: Storage And Maintenance Of The Monitor

    • You should protect the monitor from moisture, dirt, dust, blood, control solution or water, major temperature fluctuations and direct sunlight. • Do not use the monitor in the vicinity of strong electromagnetic fields and keep it away from two-way radios and mobile telephones. Beurer BGL40 mg/dL...
  • Page 43: In The Event Of Problems

    (greater than 300 mmHg) Err 04 Diastolic pressure cannot Repeat test and do not move or talk be measured during the test. Err 07 Systolic pressure out of Repeat test and do not move or talk range during the test. Beurer BGL40 mg/dL...
  • Page 44 Insert a test strip into the monitor with the contacts incompletely inserted first. Make sure that the front of the test strip is facing or wrong way around you (see "3.3 Test strips" page 16). Faulty monitor Contact Customer Service. Beurer BGL40 mg/dL...
  • Page 45 Repeat the test. Repeat test and do not move or talk during the test. Make sure your body position is correct during the test. You are not yet relaxed Rest for approx. 10 minutes. enough. Repeat the test. Beurer BGL40 mg/dL...
  • Page 46: Bgl40 Mg/Dl Technical Specifications

    11 BGL40 mg/dL TECHNICAL PECIFICATIONS 74 x 100 x 61 mm Dimensions 150 g (with batteries) Weight 2 x 1.5V AAA alkaline batteries Power supply More than 350 blood pressure tests Battery life 480 stored results with date/time Storage capacity...
  • Page 47 A test with control solution is similar to a blood test. The difference is that a drop of control solution is used instead of a drop of blood. The test result with the control solution must be within the result range. The result range is printed on every test strip container. Beurer BGL40 mg/dL...
  • Page 48 Substance for regulating viscosity 15.0% Controls The Beurer BGL40 mg/dL monitor system for private use conforms to the EU Medical Devices Directive, the German medical devices act (MPG) and European standards EN1060-1 (non-invasive blood pressure monitors, Part 1: general requirements) and EN1060-3 (non-invasive blood pressure monitors, Part 3:...
  • Page 49: Customer Service

    Customer Service address OUR OBLIGATIONS TO YOU: Our mission is to supply you with high-quality health products and the best customer service. If you are not completely satisfied with this product, please contact our Customer Service department. Beurer BGL40 mg/dL...