Beurer BM 77 - Bluetooth Upper arm blood pressure monitor Manual

Beurer BM 77 - Bluetooth Upper arm blood pressure monitor Manual

Also See for BM 77:

Advertisement

Beurer BM 77 - Bluetooth Upper arm blood pressure monitor Manual

Included in delivery

Check that the exterior of the cardboard delivery packaging is intact and make sure that all contents are present. Before use, ensure that there is no visible damage to the device or accessories and that all packaging material has been removed.
If you have any doubts, do not use the device and contact your retailer or the specified Customer Service address.
1x upper arm blood pressure monitor
1x upper arm cuff (24-40 cm)
1x instructions for use
1x brief instructions
1x storage bag
1x USB cable 1x cuff holder
4x 1.5 V AAA LR03 batteries

Signs and symbols

The following symbols are used on the device, in these instructions for use, on the packaging and on the type plate for the device:

warning
indicates a hazardous situation which, if not avoided, could result in death or serious injury.
caution
indicates a hazardous situation which, if not avoided, may result in minor or moderate injury.
information Product information
Note on important information
Observe the instructions
Read the instructions before starting work and/or operating devices or machines
Isolation of applied parts, type BF
Galvanically isolated application part (F stands for "floating"); meets the requirements for leakage currents for type BF
Direct current
The device is suitable for use with direct current only
Disposal in accordance with EC Directive WEEE (Waste Electrical and Electronic Equipment)
Do not dispose of batteries containing hazardous substances with household waste
Dispose of packaging in an environmentally friendly manner
Manufacturer
Permissible storage and transport temperature and humidity
Permissible operating temperature and humidity
Protected against foreign objects equal to or greater than 12.5 mm in diameter and against vertically falling drops of water
Serial number
CE labelling
This product satisfies the requirements of the applicable European and national directives

Intended use

Intended use
The blood pressure monitor is intended for the fully automatic, non-invasive measurement of arterial blood pressure and pulse values on the upper arm.

Target group
It is designed for self-measurement by adults in the home environment and is suitable for users whose upper arm circumference is within the range printed on the cuff. The device is also ideal for taking blood pressure measurements on women who are pregnant. This was successfully tested as part of a clinical study (Tempestas, Institut für Medizinische Forschung, Cloppenburg, Germany).

Indication/clinical benefits
The user can record their blood pressure and pulse values quickly and easily using the device. The recorded values are classified according to internationally applicable guidelines and evaluated graphically. Furthermore, the device can detect any irregular heart beats that occur during measurement and inform the user via a symbol in the display. The device saves the recorded measurements and can also output average values of previous measurements. This blood pressure monitor also has a haemodynamic stability display, which is referred to as a resting indicator throughout these instructions for use. This shows whether you, and consequently your circulatory system, are sufficiently at rest when the blood pressure measurement is being taken, and whether the measured blood pressure thus corresponds to your blood pressure when at rest. Read more about this under "Resting indicator" in the section on using the device. The recorded data can provide healthcare service providers with support during the diagnosis and treatment of blood pressure problems, and therefore plays a part in the long-term monitoring of the user's health.

Device description

Blood pressure monitor and cuff

Blood pressure monitor and cuff overview

  1. Cuff
  2. Cuff line
  3. Cuff connector
  4. Connection for mains part and USB interface
  5. Display
  6. Memory buttons M1/M2
  7. Start/stop button
  8. Risk indicator
  9. Connection for cuff connector (left-hand side)
  10. Symbol for Bluetooth® transfer
  11. Resting indicator display
  12. Cuff holder

Display

  1. Date / time
  2. Systolic pressure
  3. Diastolic pressure
  4. Pulse value
  5. Pulse symbol
  6. Release air
  7. Number of memory space/memory display average value (), morning (), evening ()
  8. Battery replacement symbol
  9. User memory
  10. Risk indicator
  11. Cardiac arrhythmia symbol
  12. Symbol for Bluetooth ® transfer

Initial use

Inserting the batteries

  • Remove the battery compartment lid on the rear of the device.
  • Insert four 1.5 V AAA (alkaline type LR03) batteries. Make sure that the batteries are inserted the correct way round.
  • Close the battery compartment lid again carefully.
  • All display elements are briefly displayed, flashes in the display. Set the date and time as described below.

If the battery replacement symbol is permanently displayed, you can no longer perform any measurements and must replace all batteries. Once the batteries have been removed from the device, the date and time must be set again. Any saved measurements are retained.

Operation with the mains part

You can alternatively also operate this device with a mains part (not included in delivery). However, before connecting the device with the mains part, please ensure that you have removed the batteries from the device. During mains operation, there must not be any batteries in the battery compartment, as this could damage the device.

  • To avoid any potential damage, the device may only be operated with a mains part that meets the specifications described in the chapter "Technical specifications".
  • Furthermore, the mains part must only be connected to the mains voltage that is specified on the type plate.
  • First, remove the protective cap of the mains part connector – located on the rear of the device – and insert the mains part into the connection intended for this purpose.
  • Then insert the mains plug of the mains part into the mains socket.
  • After using the blood pressure monitor, unplug the mains part from the mains socket first and then disconnect it from the blood pressure monitor. As soon as you unplug the mains part, the blood pressure monitor loses the date and time setting but the saved measured values are retained.

Set the hour format, date, time and Bluetooth® settings

It is essential to set the date and time. Otherwise, you will not be able to save your measured values correctly with a date and time and access them again later.
information There are two different ways to access the menu from which you can adjust the settings:

  • Before initial use and after each time you replace the battery:
    When inserting batteries into the device, you will be taken to the relevant menu automatically.
  • If the batteries have already been inserted:
    With the device switched off, press and hold the START/STOP button for approx. 5 seconds.

In this menu you can adjust the following settings in succession:

Hour format
The hour format now flashes on the display.

  • Choose the desired hour format with the M1/M2 memory buttons and confirm with the Start/stop button .

Date
The year flashes on the display.

  • Choose the desired year with the M1/M2 memory buttons and confirm with the Start/stop button .

The month flashes on the display.

  • Choose the desired month with the M1/M2 memory buttons and confirm with the Start/stop button .

The day flashes on the display.

  • Choose the desired day with the M1/M2 memory buttons and confirm with the Start/stop button .

information If the hour format is set as 12h, the day/month display sequence is reversed.

Time
The hour flashes on the display.

  • Choose the desired hour with the M1/M2 memory buttons and confirm with the Start/stop button .

The minute flashes on the display.

  • Choose the desired minute with the M1/M2 memory buttons and confirm with the Start/stop button .

Bluetooth® settings
The Bluetooth® symbol flashes on the display.

  • Use the M1/M2 memory buttons to choose whether automatic Bluetooth® data transfer is activated (Bluetooth® symbol flashes) or deactivated (Bluetooth® symbol is not shown) and confirm with the START/ STOP button .

information Bluetooth® transfers will reduce the battery life.

Usage

General rules when measuring blood pressure yourself

  • In order to generate as informative a profile of the progres-sion of your blood pressure as possible and ensure that the measured values can be compared, you should measure your blood pressure regularly and always at the same times of day.
    It is recommended that you measure your blood pressure twice a day: once in the morning after getting up and once in the evening.
  • You should always carry out the measurement when you are sufficiently physically rested. You should therefore avoid taking measurements during stressful periods.
  • Do not take a measurement within 30 minutes of eating, drinking, smoking or exercising.
  • Before the initial blood pressure measurement, make sure always to rest for 5 minutes.
  • Furthermore, if you want to take several measurements in succession, make sure always to wait for at least 1 minute between the individual measurements.
  • Repeat the measurement if you are unsure of the measured value.

Attaching the cuff

  • Fundamentally, blood pressure can be measured on both arms. Certain deviations between the measured blood pressure on the right arm and left arm are due to physiological causes and completely normal. You should always perform the measurement on the arm with the highest blood pressure values. Before starting self-measurement, consult your doctor in this regard. From this point on, always take measurements on the same arm.
  • The device may only be operated with one of the following cuffs. This should be selected in accordance with your upper arm circumference. The fit should be checked before measurement using the index mark described below.
    Ref. no. Designation Arm circumference
    163.764* Universal cuff 24-40 cm
    * Included in standard delivery
  • Place the cuff onto the bare upper arm. The circulation of the arm must not be hindered by tight clothing or similar.
  • The cuff must be placed on the upper arm so that the bottom edge is positioned 2-3 cm above the elbow and over the artery. The line should point to the centre of the palm here.
  • Guide the end of the cuff that is sticking out through the metal ring, fold it back over the arm and close the cuff using the hook-and-loop fastener. The cuff should be fastened tightly, but not too tightly, so that two fingers can still fit under the closed cuff.
  • The cuff is suitable for you if the index mark () is within the OK range after fitting the cuff.
  • Now insert the cuff line into the connection for the cuff connector.

Adopting the correct posture

  • To carry out a blood pressure measurement, make sure you are sitting upright and comfortably. Lean back and place your arm on a surface. Do not cross your legs. Place your feet next to each other flat on the floor.
  • Always make sure that the cuff is at heart level.
  • To avoid distorting the measurement, you should remain as still as possible during the measurement and not speak.

Performing the blood pressure measurement

Measurement

  • To start the blood pressure monitor, press the Start/stop button . All display elements are briefly displayed.

    The blood pressure monitor will begin the measurement automatically after approx. 3 seconds. The cuff inflates as the first step.
    information Measuring can be cancelled at any time by pressing the Start/stop button .
  • Then the air pressure in the cuff is slowly released. Now the measurement starts. As soon as a pulse is found, the pulse symbol is displayed.
  • If a tendency towards high blood pressure can already be identified, it is possible that the cuff may inflate again to a higher pressure level whilst the air is being released.
  • The remaining air is released quickly once the measurement is complete.
  • Systolic pressure, diastolic pressure and pulse measurements are displayed. On the display, the resting indicator lights up green or red depending on the assessment of your haemodynamic stability.
  • appears if the measurement could not be performed properly. In this case, please read the section "What if there are problems?".
  • Now select the desired user memory by pressing the M1 or M2 memory buttons. If you do not select a user memory, the measurement is stored in the most recently used user memory. The relevant symbol or appears on the display.
  • If the Bluetooth® function has been activated, the data transfer to the "beurer HealthManager" app starts after confirming the user selection using the START/STOP button.
  • The Bluetooth® symbol flashes on the display. The device now attempts to connect to the app for approx. 30 seconds.
  • The Bluetooth® symbol stops flashing as soon as the connection is established. All measurement data is transferred to the app. Once the data transfer is successfully complete, the device switches off automatically.
  • If a connection to the app cannot be established after 30 seconds, the Bluetooth® symbol goes out and the device switches off automatically after 3 minutes.
  • If you forget to turn off the device, it will switch off automatically after approx. 3 minutes. In this case too, the value is stored in the selected or most recently used user memory.

Transfer of measurements via Bluetooth®

In addition to displaying and saving your measurements locally on the device, you have the option of transferring your measurements to your smartphone using Bluetooth® low energy technology.
The software can be downloaded free of charge from the download area under Service at www.beurer.com.

System requirements for the " beurer HealthManager" app

  • iOS ≥ 10.0, AndroidTM ≥ 5.0
  • Bluetooth® ≥ 4.0

List of compatible devices:

To transfer the measured values, proceed as follows:

  1. BM 77
    Activate the Bluetooth® function on your device as described in section "Setting the hour format, date, time and Bluetooth®".
  2. "beurer HealthManager" app
    In the "beurer HealthManager" app, add the BM 77 under Settings/Devices and follow the instructions.
  3. BM 77
    Take a measurement.
  4. BM 77:
    • (Data transfer immediately following the measurement): If the Bluetooth® function has been activated, the data is transferred after confirming the user selection using the START/STOP button .
    • (Data transfer at a later point in time): Go into the memory recall mode for the desired user memory (see "Saving, accessing and deleting measured values"). The data transfer starts automatically.

information Also note the following information:

  • Ensure that the "beurer HealthManager" app on your smartphone is always activated and launched when you start the data transfer on the device.
  • You can tell that the data transfer is in progress by the Bluetooth® symbol shown on the display.
  • If your smartphone has a protective cover, remove it to ensure that there is no interference during the transfer.

Transfer of measurements via USB

Alternatively, you can also transfer your measured values to the PC.
To do this, you require a commercially available USB cable (included in delivery) and the "beurer HealthManager" PC software. This can be downloaded free of charge from the download area under Service at www.beurer.com.

System requirements for the „ beurer HealthManager" PC software

  • from Windows 8.1
  • from USB 2.0 (Type-A)

To transfer the measured values, proceed as follows:

  • Connect your blood pressure monitor to your PC using the USB cable provided. Please note that you cannot start a data transfer to the PC whilst a measurement is being taken.
  • PC is shown on the display. Begin the data transfer in the " beurer HealthManager" PC software.
  • During the data transfer, an animation is shown on the display. A successful data transfer is displayed as in figure 1. If the data transfer is unsuccessful, an error message appears as in figure 2. In this case, interrupt the PC connection and start the data transfer again.
  • After 30 seconds of not being in use or if communication with the PC is interrupted, the blood pressure monitor switches itself off automatically.

Evaluating the results

General information about blood pressure

  • Blood pressure is the force with which the bloodstream presses against the arterial walls. Arterial blood pressure constantly changes in the course of a cardiac cycle.
  • Blood pressure is always stated in the form of two values:
    • The highest pressure in the cycle is called systolic blood pressure. This arises when the heart muscle contracts and blood is pumped into the blood vessels.
    • The lowest is diastolic blood pressure, which is when the heart muscle has completely stretched back out and the heart fills with blood.
  • Fluctuations in blood pressure are normal. Even during repeat measurements, considerable differences between the measured values may occur. One-off or irregular measurements therefore do not provide reliable information about the actual blood pressure. Reliable assessment is only possible when you perform the measurement regularly under comparable conditions.

Cardiac arrhythmia
This device can identify potential disruptions of the heart rhythm when measuring and, if necessary, indicates this after the measurement with the symbol . This can be an indicator for arrhythmia. Arrhythmia is an illness in which the heart rhythm is abnormal because of flaws in the bioelectrical system that regulates the heartbeat. The symptoms (skipped or premature heart beats, pulse being slow or too fast) can be caused by factors such as heart disease, age, physical make-up, excess stimulants, stress or lack of sleep. Arrhythmia can only be determined through an examination by your doctor. If the symbol appears frequently, please consult your doctor. Only they are able to determine any arrythmia during an examination.

Risk indicator
The measurements can be classified and evaluated in accordance with the following table.
However, these standard values serve only as a general guideline, as the individual blood pressure varies in different people and different age groups etc.
It is important to consult your doctor regularly for advice. Your doctor will tell you your individual values for normal blood pressure as well as the value above which your blood pressure is classified as dangerous.
The bar chart on the display and the scale on the device show which category the recorded blood pressure values fall into. If the values of systole and diastole fall into two different categories (e.g. systole in the high normal category and diastole in the normal category), the graphical classification on the device always shows the higher category; for the example given this would be high normal.

Blood pressure value category Systole
(in mmHg)
Diastole
(in mmHg)
Action
Level 3:
severe hypertension
red ≥ 180 ≥ 110 Seek medical attention
Level 2:
moderate hypertension
orange 160 – 179 100 – 109 Seek medical attention
Level 1:
mild hypertension
yellow 140 – 159 90 – 99 Regular monitoring by doctor
High normal green 130 – 139 85 – 89 Regular monitoring by doctor
Normal green 120 – 129 80 – 84 Self-monitoring
Optimal green < 120 < 80 Self-monitoring

Resting indicator
One of the most common errors made when taking a blood pressure measurement is not ensuring that the user's circulatory system is sufficiently at rest when taking the measurement. In this case, the measured systolic and diastolic blood pressure values do not correspond to the blood pressure at rest. However, it is this blood pressure at rest that should be used to assess the measured values.
This blood pressure monitor uses integrated haemodynamic stability diagnostics (HSD) to measure the user's haemodynamic stability when taking the blood pressure measurement. This enables it to indicate whether the blood pressure was taken when the user's circulatory system was sufficiently at rest.

Resting indicator lights up GREEN
The measured blood pressure value was obtained when the user's circulatory system was sufficiently at rest and reliably corresponds to the blood pressure at rest.
Resting indicator lights up RED
Indicates that the value was obtained when the user's circulatory system was not sufficiently at rest. The blood pressure values measured in this case generally do not correspond to the blood pressure at rest. The measurement should therefore be repeated after a period of physical and mental rest lasting at least 5 mins.
Resting indicator does not light up During the measurement it was not possible to determine whether the user's circulatory system was sufficiently at rest. In this case too, the measurement should be repeated after a period of rest lasting at least 5 mins.

If the user's circulatory system was not sufficiently at rest, this can be caused by various factors such as physical or mental strain or distraction, speaking, or experiencing cardiac arrhythmia during the measurement.
In an overwhelming number of cases, HSD will give a very good guide as to whether the user's circulatory system is rested when taking a blood pressure measurement.
However, certain patients suffering from cardiac arrhythmia or chronic mental conditions may remain haemodynamically unstable even in the long-term, something which persists even after repeated periods of rest. The accuracy of the results for the blood pressure at rest is reduced in these users.
Like any medical measurement method, the precision of HSD is limited and can lead to incorrect results in some cases. Nevertheless, the blood pressure measurements taken when the user's circulatory system is sufficiently at rest represent particularly reliable results.

Displaying and deleting measurements

User memory
The results of every successful measurement are stored together with the date and time. If there are more than 60 measurements, the oldest measurements are lost.

  • To access memory recall mode, the blood pressure monitor must first be started. To do this press the Start/ stop button .
  • Within 3 seconds of the full-screen display appearing, select the desired user memory () with the M1 or M2 memory button.
    • To view the measurements for user memory , press the M1 memory button.
    • To view the measurements for user memory , press the M2 memory button.

Average values
If Bluetooth® is activated (the symbol appears on the display), measurements are automatically transferred. Press the M1/M2 button to cancel data transfer. The device will show average values. The symbol is no longer shown.
information
If you have selected user memory1, the M1 memory button must be pressed.
If you have selected user memory2, the M2 memory button must be pressed.

flashes on the display.
The average value of all saved measured values in this user memory is displayed.

  • Press the relevant memory button (M1 or M2).

flashes on the display.
The average value of the morning measurements for the last 7 days is displayed (morning: 5.00 a.m. – 9.00 a.m.).

  • Press the relevant memory button (M1 or M2).

flashes on the display. The average value of the evening measurements for the last 7 days is displayed (evening: 6.00 p.m. – 08.00 p.m.).

Individual measured values

  • When the relevant memory button (M1 or M2) is pressed again, the last individual measurement is displayed (in this example, measurement 03).
  • When the relevant memory button (M1 or M2) is pressed again, you can view your individual measurements.
  • To switch the device off again, press the Start/stop button or wait 30 sec.

information You can exit the menu at any time by pressing the Start/stop button .

Deleting measured values

  • To delete all measurements saved for a particular user, first select the user as described by pressing memory button M1/M2.
  • flashes on the display and the average value of all saved measured values in this user memory is displayed.
  • Depending on which user memory was selected, press and hold memory button M1 or M2 for approx. 5 seconds.
  • appears on the display. All the values in the current user memory are deleted.

Deleting Individual measurements
To clear individual measurements from the relevant user memory, you must first select a user memory.

  • Start the retrieval of the individual measured values.
  • Press and hold the memory button M1/M2 for 5 seconds (depending on the user memory you are in).
  • The selected value is deleted. The device briefly displays .
  • If you would like to delete other values, repeat the process described above.

You can switch off the device at any time by pressing the START/STOP button.

Cleaning and maintenance

  • Clean the device and cuff carefully using a slightly damp cloth only.
  • Do not use any cleaning agents or solvents.
  • Under no circumstances hold the device and cuff under water, as this can cause liquid to enter and damage the device and cuff.
  • If you store the device and cuff, do not place heavy objects on the device and cuff. Remove the batteries. The cuff line should not be bent sharply.

Accessories and replacement parts

Accessories and replacement parts are available from the corresponding service address (according to the service address list). Please state the corresponding order number.

Designation Item number and/or order number
Universal cuff (24-40 cm) 163.764
Mains part (EU) 071.95
USB cable 162.928

What if there are problems

Error
message
Possible cause Solution
You moved or spoke during the measurement. Please wait one minute and repeat the measurement.
Ensure that you do not speak or move during the measurement.
The measured blood pressure values are outside the measurement range.
The cuff was not attached correctly, the cuff line has not been inserted properly, or inflation takes longer than 15 seconds. Please wait one minute and repeat the measurement, making sure to observe the information in chapter "Attaching the cuff".
Please also check whether the cuff line has been inserted properly and ensure that neither your arm nor other heavy objects are pressing on the line, and that the line is not bent.
The inflation pressure is higher than 300 mmHg. Please take another measurement to check whether the cuff can be correctly inflated.
Make sure that neither your arm nor other heavy objects are pressing on the line, and that the line is not bent.
An error occurred when saving the measurement. Please wait one minute and repeat the measurement.
Ensure that you do not speak or move during the measurement.
Unable to transfer the data via Bluetooth®. Please observe the information in chapter "Transfer of measurements via Bluetooth®".
Unable to transfer the data to the PC. Please observe the information in chapter "Transfer of measurements via USB".
The batteries are almost empty. Insert new batteries into the device.

Specifications

Device
Model no. BM 77
Type BM 75
Measurement method Oscillometric, non-invasive blood pressure measurement on the upper arm
Measurement range Cuff pressure 0 – 299 mmHg,
systolic 30 – 260 mmHg,
diastolic 30 – 260 mmHg,
Pulse 40 –199 beats/minute
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg, pulse ± 5% of the value shown
Measurement inaccuracy Max. permissible standard deviation according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory 2 x 60 memory spaces
Dimensions L 175 mm x W 117 mm x H 50 mm
Weight Approx. 518 g (without batteries, with cuff)
Cuff size 24 to 40 cm
Permissible operating conditions +10°C to + 40°C, 15% – 85% relative air humidity (non-condensing)
Permissible storage and transport conditions -10°C to + 60°C, 10% – 90% relative air humidity, 700 –1013 hPa ambient pressure
Power supply 4 x 1,5V AAA batteries
Battery life For approx. 120 measurements, depending on the blood pressure level and/or pump pressure
Classification Internal supply, IP21, no AP or APG, continuous operation, application part type BF
Data transfer via Bluetooth ® wireless technology The blood pressure monitor uses Bluetooth® low energy technology, 2402MHz – 2480MHz frequency band, Transmission power max. 4.54 dBm, Compatible with Bluetooth® 4.0 smartphones/tablets

The serial number is located on the device or in the battery compartment.
Technical information is subject to change without notification to allow for updates.

Mains part

Model no. LXCP12-006060BEH
Input 100 – 240 V, 50 – 60 Hz, 0.5 A max
Output 6V DC, 600 mA, only in connection with beurer blood pressure monitor.
Supplier Shenzhen Iongxc power supply co., ltd
Protection This device is double insulated and protected against short circuit and overload by a primary thermal fuse.
Make sure to take the batteries out of the compartment before using the mains part.
Polarity of the the DC voltage connection
Double insulated / equipment class 2
Enclosures and Protective Covers Equipment enclosed to protect against contact with live parts, and with parts which can become live (finger, pin, hook test).
The operator shall not contact the patient and the output plug of AC mains part simultaneously.

Warnings and safety notes

warning Contraindications

  • Do not use the blood pressure monitor on newborns, children or pets.
  • People with restricted physical, sensory or mental skills should be supervised by a person responsible for their safety and receive instructions from this person on how to use the device.
  • If you have any of the following conditions, it is essential you consult your doctor before using the device: cardiac arrhythmia, circulatory problems, diabetes, pregnancy, pre-eclampsia, hypotension, chills, shaking.
  • People with pacemakers or other electrical implants should consult their doctor before using the device.
  • The blood pressure monitor must not be used in connection with a high-frequency surgical unit.
  • Do not use the cuff on people who have undergone a mastectomy.
  • Do not place the cuff over wounds as this may cause further injury.
  • Make sure that the cuff is not placed on an arm in which the arteries or veins are undergoing medical treatment, e.g. intravascular access or intravascular therapy, or an arteriovenous (AV) shunt.

warning General warnings

  • The measured values taken by you are for your information only – they are no substitute for a medical examination. Discuss the measured values with your doctor and never make your own medical decisions based on them (e.g. regarding dosages of medicines).
  • The device is only intended for the purpose described in these instructions for use. The manufacturer is not liable for damage resulting from improper or incorrect use.
  • Using the blood pressure monitor outside your home environment or whilst on the move (e.g. whilst travelling in a car, ambulance or helicopter, or whilst undertaking physical activity such as playing sport) can influence the measurement accuracy and cause incorrect measurements.
  • Cardiovascular diseases may lead to incorrect measurements or have a detrimental effect on measurement accuracy.
  • Do not use the device at the same time as other medical electrical devices (ME equipment). This could lead to a malfunction of the device and/or an inaccurate measurement.
  • Do not use the device outside of the specified storage and operating conditions. This could lead to incorrect measurements.
  • Only use the cuffs included in delivery or cuffs described in these instructions for use for the device. Using another cuff may lead to measurement inaccuracies.
  • Please note that when inflating the cuff, the functions of the limb in question may be impaired.
  • Do not perform measurements more frequently than necessary. Due to the restriction of blood flow, some bruising may occur.
  • During the blood pressure measurement, the blood circulation must not be stopped for an unnecessarily long time. If the device malfunctions remove the cuff from the arm.
  • Place the cuff on your upper arm only. Do not place the cuff on other parts of the body.
  • The air line poses a risk of strangulation for small children. Furthermore, included small parts pose a risk of suffocation for small children if swallowed. They should therefore always be supervised.

warning General precautions

  • The blood pressure monitor is made from precision and electronic components. The accuracy of the measurements and service life of the device depend on its careful handling.
  • Protect the device from impacts, humidity, dirt, marked temperature fluctuations and direct sunlight.
  • Ensure the device is at room temperature before measuring. If the measuring device has been stored close to the maximum or minimum storage and transport temperatures and is placed in an environment with a temperature of 20°C, it is recommended that you wait approx. 2 hours before using the measuring device.
  • Do not drop the device.
  • Do not use the device in the vicinity of strong electromagnetic fields and keep it away from radio systems or mobile telephones.
  • We recommend that the batteries be removed if the device is not to be used for a prolonged period of time.
  • Avoid any mechanical restriction, compression or bending of the cuff line.

Measures for handling batteries
warning

  • If your skin or eyes come into contact with battery fluid, rinse the affected areas with water and seek medical assistance.
  • Choking hazard! Small children may swallow and choke on batteries. Therefore, store batteries out of the reach of small children.
  • Risk of explosion! Do not throw batteries into a fire.
  • If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
  • Do not disassemble, open or crush the batteries.
  • Observe the plus (+) and minus (-) polarity signs.
  • Protect the batteries from excessive heat.
  • Do not charge or short-circuit batteries.
  • If the device is not to be used for a relatively long period, take the batteries out of the battery compartment.
  • Use identical or equivalent battery types only.
  • Always replace all batteries at the same time.
  • Do not use rechargeable batteries!

warning Notes on electromagnetic compatibility

  • The device is suitable for use in all environments listed in these instructions for use, including domestic environments.
  • The use of the device may be limited in the presence of electromagnetic disturbances. This could result in issues such as error messages or the failure of the display/device.
  • Avoid using this device directly next to other devices or stacked on top of other devices, as this could lead to faulty operation. If, however, it is necessary to use the device in the manner stated, this device as well as the other devices must be monitored to ensure they are working properly.
  • The use of accessories other than those specified or provided by the manufacturer of this device can lead to an increase in electromagnetic emissions or a decrease in the device's electromagnetic immunity; this can result in faulty operation.
  • Failure to comply with the above can impair the performance of the device.

Documents / Resources

References

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.