If the results are outside the specified range, check the following possible causes:
Cause
The code number on the display is not
the same as that of the test strip
container.
• You did not dispose of the first drop
of control solution
• The vial tip was not wiped clean.
• The vial was not shaken sufficiently.
Control solution has expired or is
contaminated.
Control solution or monitor is too warm
or too cold.
Damaged test strips
Old test strips
A problem with the monitor
WARNING
If you receive repeated results outside the specified range with the
control solution, do not continue using the system to measure your
blood glucose level. Contact our Customer Service department.
Beurer BGL40 mg/dL
Measure
Code the monitor (see "6.1 Checking
code number (coding)" page 24).
Correct the cause and repeat the
test.
Repeat the test with a new vial of
control solution.
Bring control solution and monitor to
room temperature (+20°C to +25°C)
and repeat test.
Repeat the test with a new test strip.
Open a new test strip container.
Repeat the test.
Contact
our
department.
Customer