6
2.3

Error output

The RLI20 outputs on RS422 compatible error signal.
Cabling
A screened cable is recommended. The connector on the RLI20
is a 4 way female Binder M9, the mating connector is shown in
Figure 3. Pinouts are shown in Table 4.
Table 4 – Error connector pinout
Pin
1
2
3
4
Shell
Figure 3 – 4 way male Binder
Function
/Error
0 V
+5 V
Error
Chassis
2.4

Power

The RLI20 power is supplied via a 4 way male M16 Binder,
the mating connector is shown in Figure 4. Pinouts are shown
in Table 5. This is identical to wiring requirements for RCU10
power connector.
Table 5 – Power connector pinout
Pin
1
2
3
4
Case
The 24 V power supply should be single fault tolerant certified to
EN (IEC) 60950-1.
Figure 4 – 4 way female M16 Binder connector
Installation
Function
24 V
24 V
0 V
Chassis
Chassis