Advertisement

5-4. Built-in calibration sensor and SelfCalibration problems

Problem
1. The built-in calibration sensor does not
come out/stays out.
2. SelfCalibration cannot be executed.
3. SelfCalibration failure
Error Code Table
Error Code
0011
• Calibration cannot be performed because the target brightness is too low. Change the target value, and execute SelfCalibration again.
0013
• Light may have entered the built-in calibration sensor during measurement. Ensure that light does not enter the sensor, and then execute
0014
• The target brightness may be too high. Change the target value, and then execute SelfCalibration again.
• The measurement accuracy of the built-in calibration sensor may be low. Try executing correlation using ColorNavigator.
• Adjust using ColorNavigator.
0030
• Communication with the built-in calibration sensor failed. Turn off the main power, wait for a few minutes before turning the power back on, and
0032
0041
0033
• The built-in calibration sensor may not have worked correctly. Turn off the main power, wait for a few minutes before turning the power back on,
0034
• The built-in calibration sensor may not have come out during calibration, or external light may have entered the sensor during measurement.
0036
• SelfCalibration may not have worked correctly. Check the target value, and then execute SelfCalibration again.
0037
• The built-in calibration sensor may not have worked correctly. Try executing correlation using ColorNavigator.
0038
• The built-in calibration sensor may not have worked correctly. Turn off the main power, wait for a few minutes before turning the power back on,
0039
0040
• Turn off the main power, and then turn it on again a few minutes later.
• Check whether the date and time are set correctly on the monitor (see
• Check whether the execution schedule been set (see
• Check whether the calibration targets are set correctly (see
• Set the SelfCalibration target using ColorNavigator.
• Refer to the error code table. If an error code that does not appear in the error code table is displayed, contact your local
EIZO representative.
SelfCalibration again. The built-in calibration sensor may not have worked correctly. Try executing correlation using ColorNavigator.
then execute SelfCalibration again.
and then execute SelfCalibration again.
Turn off the main power, wait for a few minutes before turning the power back on, and then execute SelfCalibration again.
and then execute SelfCalibration again.
Possible cause and remedy
"Schedule" (page
"Target" (page
Description
- 44 -
Chapter 5 Troubleshooting
"Clock Adjustment" (page
30)).
28)).
25)).