Error Code
0050
• Failed to initialize the built-in calibration sensor. External light may have entered the built-in calibration sensor. Ensure that light does not enter
the sensor, and then execute SelfCalibration again.
• Communication with the built-in calibration sensor may have failed. Turn off the main power, wait for a few minutes before turning the power
back on, and then execute SelfCalibration again.
0060
• The built-in calibration sensor may not have worked correctly.
0061
• Check whether there is any foreign object near the built-in calibration sensor.
• Execute SelfCalibration again.
Description
- 45 -
Chapter 5 Troubleshooting