Specifications

Performance

Saturation (SpO
75% ±1 Sat Point
90% ±1 Sat Point

Pulse Rate

60 bpm ±1 bpm
200 bpm ±2 bpm

Light Levels

Low Level = 750 nAi ±10%
High Level = 3,000 nAi ±10%

Modulation (%MOD)

Low Modulation = 0.4% to 0.6%
High Modulation = 4.5% to 5.5%

ECG

Analog Outputs (R-Wave)
Output 1: = RA (White)
Output 2: = LA (Black)
Output 3: = LL (Red)
This is a 3 lead ECG compatible output.