in the health-care environment (for example, cellular phones,
mobile two-way radios, electrical appliances), it is possible
that high levels of such interference due to close proximity or
strength of a source, may result in disruption of performance
of this device.
This equipment generates, uses and can radiate radio
frequency energy and if not used in accordance with the
instructions, may cause harmful interference with other
devices in the vicinity. Disruption or interference may be
evidenced by erratic readings, cessation of operation, or other
incorrect functioning. If this occurs, the site of use should be
surveyed to determine the source of this disruption, and
actions taken to eliminate the source.
The user is encouraged to try to correct the interference by
one or more of the following measures:
SRC-MAX
Turn equipment in the vicinity off and on to isolate
the offending equipment
Reorient or relocate the other receiving device
Increase the separation between the interfering
equipment and this equipment
If assistance is required, contact Nellcor Puritan
Bennett's Technical Services Department or your
local Nellcor Puritan Bennett representative.
Troubleshooting
33