Testing The Red Led - Nellcor OxiMax SRC-MAX Technical Manual

Pulse oximetry functional tester
Table of Contents

Advertisement

Testing the RED LED

1. Turn on the signal generator.
2. Adjust the output until there is 475 to 525 millivolts peak
3. Connect one end of a 1 meg ohm resistor (R2 in Figure
4. Connect the other end to the positive terminal of the
5. Connect the negative terminal of the +5 volt supply to
Caution: See Figure 11. Prior to connecting the
oscilloscope's negative terminal to the +5 volt supply (+)
terminal, ensure that the oscilloscope negative is isolated
from the +5 volt supply ground.
6. Connect the oscilloscope channel 2 (+) lead to the
7. Connect the oscilloscope channel 2 (-) lead to the
8. With Low-Light setting selected on the SRC-MAX,
SRC-MAX
across the 10 ohm resistor.
11) to pin 9 of the 9 pin D-sub female connector.
power supply set for +5 volts.
pin 5 of the 9 pin D-sub female connector.
junction of pin 9 and the 1 meg ohm resistor.
junction of the positive terminal of the power supply (set
for +5 volts) and the other end of the 1 meg ohms.
measure the peak-to-peak voltage across the 1 meg ohm
resistor on channel 2 of the oscilloscope. the peak-to-
peak voltage should be 600 to 900 millivolts peak-to-
peak. See Figure 12. The peak-to-peak voltage tolerance
includes a possible probe impedance error.
Performance Verification
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents