Teledyne 6200A Manual page 122

Trs analyzer
Table of Contents

Advertisement

TAI Model 6200A SO
Table 7-11: EPA Expected Span Gas Concentration Procedure
Step Number
Action
1.
Press CAL-
CONC
2.
Press ENTR
3.
Press EXIT
Sample the generated concentration until the SO
Span the instrument by the following procedure:
Analyzer Operator Manual, 02164, Rev. G
2
Figure 7-2: Diagram Of Calibration System
Comment
This key sequence causes the 6200A to prompt for the
expected SO
Enter the SO
under each digit until the expected value is set.
ENTR stores the expected SO
Returns instrument to SAMPLE mode.
7-15
concentration.
2
span concentration value by pressing the key
2
span value.
2
response is stabilized.