TAI Model 6200A SO
Table 6-2: Setup Automatic Zero/Span Calibration
Timer Enable
Starting Date
Starting Time
Delta Days
Delta Time
Duration
Calibrate
The Timer Enable can be set to "OFF" to disable the automatic calibration timer while the
remote RS-232 calibration of specific sequence can be initiated.
Avoid setting two or more sequences at the same time of the day. Any new
sequence which is initiated whether from a timer, the RS-232, or the contact
closure inputs will override any sequence which is in progress.
The programmed start time must be a minimum of
Examples of possible sequences are as following under any one of three available SEQx.
Example 1: to perform zero-span calibration check once per day at 10:30 pm, 12/20/98.
1) MODE: ZERO-HI
2) TIMER ENABLE: ON
3) STARTING DATE: 12/20/98
4) STARTING TIME: 22:30
5) DELTA DAYS: 1
6) DELTA TIME: 00:00
7) DURATION: 15.0 MINUTES
8) CALIBRATE: OFF
Analyzer Operator Manual, 02164, Rev. G
2
Default
Description
ON
Enable or disable automatic calibration timer.
01-JAN-95
MM:DD:YY
00:00
HH:MM, 0 - 23 hours and 0 - 59 minutes
1
Delay days between each calibration (0 - 365 days)
00:00
HH:MM, 0 - 23 hours and 0 - 59 minutes
15.0 minutes
1 - 60.0 minutes
OFF
on/off. If ON is selected, it will adjust the calibration.
NOTE
5 minutes later than the real time clock.
6-4