3.2 EPA Equivalency Designation

Teledyne Analytical Instruments, Model 6200A Sulfur Dioxide Analyzer is designated as Reference
Method Number EQSA-0495-100 as defined in 40 CFR, Part 53, when operated under the following
conditions:
Range: Any range from 50 ppb to 1000 ppb.
1. Ambient temperature range of 5 to 40
2. Line voltage range of 105-125 VAC, 60Hz.
3. With 5-micron TFE filter element installed in the internal filter assembly.
4. Sample flow of 700 +/- 100%.
5. Vacuum pump (internal or external) capable of 14"Hg Absolute pressure @ 1 slpm or better.
6. Software settings:
A. Dynamic span
B. Dynamic zero
C. AutoCal
D. Dual range
E. Autorange
F. Temp/Pressure compensation
Under the designation, the Analyzer may be operated with or without the following options:
1. Rack mount with chassis slides (P/N 01469A).
2. Rack mount without slides, ears only (P/N 01470A).
3. Fluorocarbon zero/span valves (P/N 01491A).
4. Internal zero/span (P/N 01441A).
5. SO
Permeation tube - uncertified 0.4ppm @ 0.7 lpm (P/N 0150603A)
2
6. SO
Permeation tube - certified 0.4ppm @ 0.7 lpm (P/N 0150604A)
2
7. SO
Permeation tube - certified 0.8ppm @ 0.7 lpm (P/N 0150607A)
2
8. SO
Permeation tube - uncertified 0.8ppm @ 0.7 lpm (P/N 0150608A)
2
9. Zero air scrubber (P/N 01440A).
4-20mA, isolated outputs (P/N 01471-1A).
Analyzer Operator Manual, 02164, Rev. G