Das Commands And Reports - Teledyne 6200A Manual

Trs analyzer
Table of Contents

Advertisement

TAI Model 6200A SO
The following is a summary of the Diagnostic commands.
Table 6-14: RS-232 Diagnostic Command Summary
Command
D LIST
D name=value
D LIST NAMES
D ENTER SIG
D ENTER OT
D ENTER ET
D ENTER TASK
D EXIT
D RESET
D RESET RAM
D RESET EEPROM

6.8.7 DAS Commands and Reports

RS-232 Commands
In addition to accessing the data acquisition system and the stored data from the instrument front
panel, you can also access the data acquisition and the stored data from the RS-232 interface.
There are two RS-232 commands, listed in the table below.
Table 6-15: RS-232 DAS Commands
Command
D [id] PRINT ["name"]
D [id] REPORT "name" [RECORDS=number]
[COMPACT|VERBOSE]
Analyzer Operator Manual, 02164, Rev. G
2
Description
Prints all I/O signal values. See Table 9-4 for Sig I/O definitions.
Examines or sets I/O signal. For a list of signal names see Table 9-4
in Section 9. Must issue D ENTER SIG command before using this
command.
Prints names of all diagnostic tests
Executes SIGNAL I/O diagnostic test.
Executes Optic Test diagnostic test.
Executes Elect Test diagnostic test.
Displays a listing of the tasks and their status.
Use D EXIT to leave these diagnostic modes.
Must use this command to exit SIG, ET or OT Diagnostic modes
Resets analyzer software. (Same as power on.)
Resets analyzer software and erases RAM. Erases SO
Keeps setup variables and calibration. (Same as installing new
software version.)
Resets analyzer software and erases RAM and EEPROM. Returns all
setup variables to factory defaults, resets calibration value.
RS-232 DAS Commands
Description
Prints specified data channel's properties
Prints the data for the specified data channel
6-21
conc values.
2