Other Measurements

Measuring Gaps in Sensing Objects
Gaps (G) in sensing objects can be measured using the thickness measurement.
The procedure is the same one used to measure thickness.
Measuring Thickness, p. 52
Gap Measurement Basics
Using the Scale Inversion Function
Switch ON the Scale Inversion Function (INV) on the Amplifier Units to match changes
in gap size to changes in measured values on the display. (Inversion is turned OFF in
the default setting.)
For the default setting (OFF), the display value increases when the Sensor Head plunger is pressed.
If a measurement is taken with this setting, the display value will increase as the gap size decreases.
Changes are inverted if the Scale Inversion Function (INV) is set to ON. This means that changes in
the amount of displacement can be matched to changes in display values.
Setting the Gap (G) on the Zero Reset Display (ZRDSP)
MAIN APPLICATIONS AND SETTING METHODS
Small gap = Higher display value
CH1
G
CH2
G
Large gap = Lower display value
Set the reference sample gap.