Correcting the Distance Display (Span Adjustment)

This setting is used to correct discrepancies between the display value and the actual amount
of displacement. Set the desired distance (units: µ m).
This correction does not change the Sensor's reference position (distance display: 0 mm).
The correction range is about 2% of the measurement distance.
Example: ZX-TDS01T
Measurement distance: 1 mm, range of correction: 20 µ m
Example: Actual displacement: 0.5 mm, display value: 0.498 mm
The displacement discrepancy ranges from 0 to 0.5 mm and is 0.002 mm here.
In a range from 0 to − 0.5 mm, a discrepancy of 0.002 mm is about what you would expect. If the total
discrepancy is 0.004 mm (4.0 µ m), then set the correction to 4.0.
Sensor Head
Sensing
object
Calipers
Correction details are registered in the Sensor Head. This means that once done, the correction will not have
to be redone even if the Amplifier Unit is replaced.
Moving to FUN and SPCL
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT Keys to display
SPCL on the main display.
Actual displacement
:0.5mm
Display value
:0.498mm

AUXILIARY FUNCTIONS

Display value
(mm)
0.500
0.498
Actual displacement
0
0.5
POWER
Section 6
(mm)
ZERO
ENABLE
(mm)