Settings for High-Speed Digital Output

Of CompoWay I/F or non-procedural communication commands, if you want to use the
FlowDATA command to batch acquire data at high speed, set this setting to ON.
Note, however, that in measurement modes ([High] or [Custom]) having a short sampling
cycle, the sampling cycle changes according to the setting of this digital output.
• This setting automatically is set to ON when the graph display and logging functions of SmartMonitor
ZS are used. For example, if the mode is set to High-Speed mode at this time, the sampling cycle
becomes 1.5 times the sampling cycle.
• When a measurement mode other than [High] or [Custom] is selected, this digital output setting does
not cause the sampling cycle to change.
FUN mode-[I/O SET]-[DIGITAL]
Setting
ON
(default value)
OFF
In Multi-Task mode, set ON/OFF for each TASK.
Of the communication commands, the command (FlowDATA) for high-speed batch
acquisition of data is enabled.
Set this to ON to batch acquire data at high speed using communication com-
mands.
Note, however, that in measurement modes ([High] or [Custom]) having a short
sampling cycle, the sampling cycle becomes 1.5 times the sampling cycle when
this setting is set to ON.
Of the communication commands, the command (FlowDATA) for high-speed batch
acquisition of data is disabled.
When [High] is selected for the measurement mode, this setting automatically
turns OFF, and the sampling cycle becomes the fastest sampling cycle.
Settings for High-Speed Digital Output
Description
Section7
ZS-HL
7-7
User's Manual

Hide quick links: