Setting Focus

Enter the measured values for any two binary output values to set the output range.
(default value: OFF)
Example: Set 0 mm to 0 and 5 mm to 50000. (when ZS-HLDS5 is connected)
Output value
65535
52768
32767
12768
0
-5
0
5
Separate the two specified points by at least 1 % of the rated measurement range for the connected
Sensor Head.
Example: For the ZS-HLDS5T
The two measured points must be separated by at least "10 mm × 0.01 = 0.1 mm" as the measuring
range is "±5 mm (10 mm)".
FUN mode-[I/O SET]-[TERMINAL]-[FOCUS]-[ON]
1.
Set the binary output value of point 1.
The output value is displayed on the main
display.
2.
Press the SET Key to apply the setting.
3.
Set the measured value for point 1.
The measured value is displayed on the sub-
display.
4.
Press the SET Key to apply the setting.
Output value
65535
50000
32767
Measured value
0
(mm)
Settings for Terminal Block Output
Measured value
(mm)
-5
0
5
INPUT POINT 1
DIG
VAL SET:OK
INPUT POINT 1
DIG
VAL SET:OK
Section6
ZS-HL
6-13
User's Manual