UNI-T UT8806E User Manual page 47

Desktop digital multimeter
Hide thumbs Also See for UT8806E:
Table of Contents

Advertisement

User's Manual
Figure 0–29 Measuring Connection of Capacitance Measurement
3. Select the appropriate range according to the size of the capacitance to be
The user can press [Range] to select a measuring range or use the front panel keys
quickly select a range. The automatic range automatically selects a range for the
measurement based on the input. Compared to the manual range, the automatic range is
more convenient, but the measurement speed is slower.
Table 0-9
Scale*
Input
Protection
Configuration
Parameter
Notes*
 All ranges are allowed to over range by 20%.
 "OL" will be displayed when the maximum range.
4.
Set the relative value
Turn on or off the relative operation function. When the relative operation is turned on,
the displayed number is the actual measured value minus the set relative value. (For
details on how to set the relative value, refer to the section "Mathematical Operation
Function" in this chapter).
5.
Mathematical operation
The user can perform mathematical operations (statistic, limit, relative) on DCI
measurement readings. For details on how to use this function, refer to the section
"Mathematical Operation Function" in this chapter.
Instruments.uni-trend.com
2nF, 20nF, 200nF, 2uF, 20uF, 200uF, 2mF, 20mF, 100mF
All scales are DC 1000V or AC 750Vrms (Hi)
Range, the set value of relative operation
UT8806E Desktop Digital Multimeter
measured.
to
36 / 104

Advertisement

Table of Contents
loading

Table of Contents