Input Impedance - UNI-T UT8806E User Manual

Desktop digital multimeter
Hide thumbs Also See for UT8806E:
Table of Contents

Advertisement

User's Manual
4. For CAP measurement, the resolution is fixed at
5. For CONT measurement, it always displayed to 2 decimal places.
6. For DIODE measurement, the resolution is fixed at
7. For TEMP measurement, the resolution is fixed at
RTD 2W、 RTD 4W
Thermis 2W、 Thermis 4W
Thermocouple (J、K、E、T、N)
Thermocouple (R、S、B)

Input Impedance

The input impedance setting applies to the DCV measurement function. The factory default is
"Auto". For 200mV, 2V and 20V scales, "10 MΩ" can be selected to minimize the load error
introduced by the multimeter to the measured object.
Currently, the measurement function is DCV, and the range is Auto or 10MΩ. Press the Menu
key to set the [Input Impedance] in the menu, as shown in Figure 2-4.
Press [Auto] or [10MΩ] to select the impedance value.
 Select "Auto", then the input impedance of all ranges is greater than 10 GΩ.
 Select "10MΩ", then the input impedance of all ranges is equal to 10 MΩ.
Instruments.uni-trend.com
Always display to 3 decimal places
Always display to 2 decimal places
Always display to 1 decimal places
Figure 0–4 Select DC Impendence
UT8806E Desktop Digital Multimeter
bit.
bit.
bit.
15 / 104