UNI-T UT8806E User Manual page 37

Desktop digital multimeter
Hide thumbs Also See for UT8806E:
Table of Contents

Advertisement

User's Manual
Table 0-5
4. Set the filter
Press the [Filter] to configure an appropriate filter for measurement. The UT8806E
provides three kinds of filter, ">3Hz, >20Hz and >200Hz". Normally, the user should select
the highest frequency filter whose frequency is less than the frequency of the
measurement signal.
5. Set the relative value
Turn on or off the relative operation function. When the relative operation is turned on, the
displayed number is the actual measured value minus the set relative value. (For details on
how to set the relative value, refer to the section "Mathematical Operation Function" in this
chapter).
6. Read the measured results
When reading the measured results, if press the
can get the frequency of measured signal, as shown in Figure 2-16.
Instruments.uni-trend.com
200mV,2V,20V,200V,750V
Scale*
Input
All scales are DC 1000V or AC 750Vrms
Protection
(Hi end)
Configuration
Range, filter, reading rate, the set
Parameter
value of relative operation
Notes*
 Except 750V, all ranges are allowed to over range
by 20%.
 "OL" will be displayed when the input exceeds 787V
in 750V range.
 Input protection of 750Vrms is for every range.
UT8806E Desktop Digital Multimeter
key, set [Dual] to "Frequency", it
26 / 104