Installing / removing and connecting analyzer and option modules
7.5 Installing OXYMAT 7, high-temperature version
7.5
Installing OXYMAT 7, high-temperature version
Overview
①
②
③
Figure 7-4
Procedure
1. Isolate the wall-mounted device from power.
2. Loosen the six screws
3. Remove the cables for equipotential bonding
Note: The cable for equipotential bonding
module
70
Wall-mounted device
Screw for door (6 units)
Torx screw for housing floor (8 units)
Installing OXYMAT 7 high-temperature
②
⑥
is installed. See section "Installing analyzer modules, standard version (Page 68)".
④
Flat seal
⑤
Equipotential bonding cable
⑥
Analyzer module OXYMAT 7, high-temperature ver‐
sion
and open the door.
⑤
⑤
is no longer required after the analyzer
at the blank plates of the two slots.
Operating Instructions, 07/2023, A5E31930403-AB
Wall-mounted device