Siemens 7MB3000 Series Operating Instructions Manual page 168

Wall-mounted device
Table of Contents

Advertisement

Technical specifications
13.3 Technical specifications of analyzer modules
Table 13-18
ULTRAMAT 7: Measuring ranges
Measuring ranges
Number of meas‐
Max. 4; parameters can be assigned freely
uring ranges
 
(combination with O7)
Table 13-19
ULTRAMAT 7: Gas inlet conditions
Gas inlet conditions
Sample gas pressure Normal pressure (atmospheric pressure
compensation)
Pressure loss between sample gas inlet and sample gas outlet
Sample gas flow
ULTRAMAT 7 standard version
ULTRAMAT 7 high-temperature
Sample gas temperature and humidity
ULTRAMAT 7 standard version
ULTRAMAT 7 high-temperature
Nominal values in the section "Determining the technical specifications"
1)
Wall-mounted device only
2)
168
Component
Smallest possible measuring span
CO
CO
2
CH
4
C
H
2
4
SO
2
NO
N
O
2
NH
3
CO/NO
CO/NO
CO 0 ... 100/NO 0 ... 300 ppm
CO
/CO
2
0 ... 10 ppm
0 ... 5 ppm
0 ... 50 ppm
0 ... 300 ppm
0 ... 50 ppm
0 ... 100 ppm
0 ... 50 ppm
0 ... 100 ppm
0 ... 100 ppm
0 ... 100 ppm
500 to 1 500 hPa (absolute)
< 10 hPa at 1.5 l/min
18 ... 90 l/h (0.3 ... 1.5 l/min) 
36 .. 90 l/h (0.6 ... 1.5 l/min)
0 to 50 °C (32 ... 122 °F) at relative humidity < 90%, in
reference to the respective ambient temperature; avoid
condensation in gas path (moisture content results in
measurement errors)
0 to 90 °C (32 ... 194 °F), avoid condensation in the gas path
(moisture content results in measurement errors)
Operating Instructions, 07/2023, A5E31930403-AB
Largest possible measuring span
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 100%
0 ... 3%
0 ... 100%
0 ... 100%
0 ... 10000 ppm
CO 0 ... 3000 ppm/NO 0 ... 3000 ppm
0 ... 100%
1, 2)
1, 2)
Wall-mounted device

Advertisement

Table of Contents
loading

Table of Contents