Note
Device variant Cleaned for O
In the device version with the order suffix -B06 (Cleaned for O
gas are cleaned and absolutely grease-free.
NOTICE
Formation of condensation in the device
Condensation can be caused by a high difference in temperature between transportation and
storage and the installation location.
• Before taking the device into operation let the device adapt for several hours in the new
9.2
General information on commissioning
Note
Ambient temperature
Ensure that the ambient temperature is within the permissible range during operation. The
corresponding information is available in the section "Technical specifications".
The technical specifications of the device can deviate outside these temperature limits.
Operability of the device is no longer ensured at deviations of more than 5 K.
Note
Vibrations at the installation site disturb the measured signal
When you use OXYMAT 7 and/or ULTRAMAT 7 analyzer modules, vibrations, especially in the
frequency range from 6 Hz to 15 Hz, can influence the measured signal.
• Select an installation location which is vibration-free.
• If necessary, change the measuring frequency settings:
– Make sure that the measuring frequencies do not correspond to the fundamental of the
– You must also ensure that the measuring frequency differs from half the basic frequency
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB
environment.
installation site nor one of its harmonic multiples.
of the vibration.
9.2 General information on commissioning
2
Commissioning
), all parts wetted by the sample
2
123