Operation With At Least One Ultramat 7; Operation With At Least One Calomat 7 - Siemens 7MB3000 Series Operating Instructions Manual

Wall-mounted device
Table of Contents

Advertisement

Table 5-1
Operating
mode
Pressure op‐
eration
Suction oper‐
ation
5.4.4

Operation with at least one ULTRAMAT 7

Table 5-2
Operating mode
Pressure operation/Suction operation
5.4.5

Operation with at least one CALOMAT 7

Table 5-3
Operating mode
Pressure operation/Suction operation
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB
Mounting position of the sample gas restrictor: OXYMAT 7
Sample gas restrictor/clamping screw
Operation with one
OXYMAT 7
Install a sample gas restric‐
tor in front of the sample
gas inlet.
Install a sample gas restric‐
tor in the sample gas outlet
Mounting position for the sample gas restrictor: ULTRAMAT 7
Mounting position for the sample gas restrictor: CALOMAT 7
Operation with two OXY‐
MAT 7 in series
Install a sample gas restric‐
tor in front of the sample
gas inlet into which the sam‐
ple gas is introduced first.
In addition replace the sam‐
ple gas restrictor in the inlet
of the rear, second module
with a clamping screw.
Install a sample gas restric‐
tor in the sample gas outlet
from which the sample gas
is first sucked out:
In addition replace the sam‐
ple gas restrictor in the inlet
of the rear, second module
with a clamping screw.
Sample gas restrictor/clamping screw
Operation with one or two ULTRAMAT 7
No sample gas restrictor/clamping screw required Subse‐
quent upgrading or retrofitting is not possible and not re‐
quired.
There are no particular mounting restrictions.
Clamping screw
Operation with one or two CALOMAT 7
Pre-assembled clamping screws. No retrofitting or disman‐
tling required.
Combined operation
5.4 Application planning
Operation with two OXY‐
MAT 7 in parallel
Install a sample gas restric‐
tor in front of each sample
gas inlet.
Install a sample gas restric‐
tor at each sample gas out‐
let
49

Advertisement

Table of Contents
loading