Technical Specifications Of Analyzer Modules; Oxymat 7 Technical Specifications - Siemens 7MB3000 Series Operating Instructions Manual

Wall-mounted device
Table of Contents

Advertisement

Climatic conditions
Ambient humidity (rel. humidity)
during storage, transport or opera‐
tion
Permissible ambient temperature
and humidity
13.3

Technical specifications of analyzer modules

13.3.1

OXYMAT 7 technical specifications

Table 13-7
OXYMAT 7: General technical specifications
General
Information on the smallest span
Max. power consumption
Weight (only module)
Electrical safety
Max. EMC error during operation with one and/or two
OXYMAT 7 / ULTRAMAT 7 units
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
1)
Table 13-8
OXYMAT 7: Measuring ranges
Measuring ranges
Number of measuring ranges
Parameters can be as‐
signed in the measur‐
ing ranges
The high-temperature version of the analyzer module is not intended for installation in the rack-mounted device.
1)
Note
Temperature change for high-temperature OXYMAT 7
If you change the temperature range, the device has to be re-calibrated. Detailed information
can be found in the operating manuals, see "Bibliography" (Page 179).
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB
< 90%
Technical specifications of analyzer modules (Page 163)
Warm-up phase
Smallest possible measuring
span
Largest possible span
13.3 Technical specifications of analyzer modules
OXYMAT 7
See module nameplate or device certificate
≤ 60 W
< 5.5 kg
In accordance with DIN EN 61010-1
 
OXYMAT 7
Max. 4; parameters can be as‐
signed freely
0.5% O
2
100% O
2
Technical specifications
OXYMAT 7 high-tempera‐
ture
1)
≤ 120 W
< 7.5 kg
1.0% according to EN
61326-1 (2013)
OXYMAT 7 high-temperature
Max. 4; parameters can be as‐
signed freely
2% O
2
100% O
2
1)
163