Commissioning
9.1 General safety instructions for commissioning
Leaking of toxic gases when the device is vented
Danger of poisoning
Toxic sample gases and liquids can be released when the device is vented.
• Before venting, ensure that there are no toxic gases or liquids in the device.
Danger of poisoning by escaping gas
Leaky gas paths lead to accumulation of the sample gas in the device.
• Check whether the threaded joints at the gas connections are secure. Tighten the threaded
• Check the gas paths for leaks. See section "Checking gas paths for leaks (Page 133)".
Hot surfaces resulting from hot sample gases and heated devices
Danger of burns resulting from surface temperatures above 70 °C (155 °F).
• Take appropriate protective measures, for example contact protection.
• Make sure that the maximum ambient temperature is not exceeded by implementing
Ignition hazard
Because of the danger of ignition, all parts coming into contact with oxygen must be clean. This
means they must be free of loose parts or parts which could become loose during operation
and of foreign particles, especially oil, grease and solvents.
Observe the following when working on analyzers with specially cleaned gas path:
• No clothing which is contaminated by oil or grease may be worn.
• The installation site must be clean and dust-free.
• Only parts which are packed and marked accordingly may be assembled.
• Always wash your hands before commencing work.
• The cleaned parts must only be handled on the surfaces which do not come into contact
• New filters, new connecting hoses, and oil-free mixing equipment must always be used to
122
WARNING
WARNING
joints with a suitable open-ended wrench, if necessary. In the process, make sure that you
counter properly.
CAUTION
protective measures. Refer to the information in the section "Technical specifications
(Page 161)".
WARNING
with the sample gas.
check the analyzer function.