Installing / removing and connecting analyzer and option modules
7.8 Removing analyzer modules, standard version
7.8

Removing analyzer modules, standard version

Overview
Figure 7-9
Procedure
1. Isolate the wall-mounted device from power.
2. Loosen the six screws
76
Wall-mounted device
Screw for door (6 units)
Flat seal
Removing an analyzer module, example OXYMAT 7
Equipotential bonding cable
Torx screw for housing floor (8 units)
OXYMAT 7 analyzer module
and open the door.
Operating Instructions, 07/2023, A5E31930403-AB