Figure 5-14 Panel Mounting Assembly - Omega DP461-S User Manual

High performance strain gage indicator
Table of Contents

Advertisement

5
Setup
METER
6. Punch or cut a hole in the panel using the panel cutout dimensions in
7. Insert the panel-mount gasket around the rear of the case and slide it
8. Working from the front of the panel, insert the case assembly, rear end
9. Working from the rear of the panel, slide the sleeve forward over the
The panel should now be sandwiched between the bezel-backed gasket in
front and the sleeve in back.
10. Replace the thumbnuts that secure the sleeve tabs to the case.
5-12
PANEL THICKNESS
0.25 [6.4] MAX
0.03 [0.8] MIN
0.06
R
[1.5]
1.772 +0.024/-.000
4PLCS
[45.00 +0.61/-0.00]
3.622 +0.032/-.000
[92.00 +0.81/-0.00]
PANEL
GASKET
BEZEL
Figure 5-14. Panel Mounting Assembly
Figure 5-14. Remove burrs and paint the panel as required.
forward to the bezel (if it's not already in place).
first, all the way into the panel cutout such that the gasket firmly backs
the panel surface.
case and up to the panel surface.
THUMBNUTS
CASE
REAR
PROTECTIVE
COVER
COVER
MOUNTING
SCREW
SLEEVE

Advertisement