9
Meter Function Menus

9.2.3 RDG SC (Reading Scale) And RDG OF (Reading Offset)

Typically, this would be used only when you cannot connect a known load to
your meter, you require a display with 3 or more positions to the right of the
decimal point position, you have scaled your meter using Input Scale and
Offset ("IN.SC.OF") and want to enter a constant multiplying factor, or you
have an extremely large offset.
If "RDG.1=0" were chosen, then you go automatically into "RDG SC" and
"RDG OF" where:
"RDG SC" is reading scale from -99999 to +499999 where you set the
display to "1.00000" and "RDG OF" is reading offset from -99999 to 999999
when you set the display to "000000".
When "RDG SC" is displayed, press the 'MIN' button to see the previously-
set value. Process measurement scale is set to "1.00000" using the
'MIN/MAX' buttons. Store by pressing the 'MENU' button.
For "RDG OF", you may choose to enter a reference temperature offset here
(e.g., "-100.00") so that the display will read deviation of the input from the
boiling point (or some other temperature).
If "RDG.1 = 1" were chosen, then you go automatically into "RD.SC.OF".
Direct reading offset "RDG OF" now has programmable decimal point
Note
position. This makes the decimal point independent of the display decimal
point and allows larger offset values.