Omega DP461-S User Manual page 102

High performance strain gage indicator
Table of Contents

Advertisement

9
Meter Function Menus
MENU
BUTTON
MAIN MENU
"L4 CNF":
* The 'MIN' button allows you to sequence through L4C.1, L4C.2, L4C.3,
L4C.4, L4C.5 and L4C.6.
The 'MAX' button allows you to select the "0" or "1" state for each "L4C"
condition.
The 'MENU' button stores the selected values for all "L4C" condition(s)
changed and advances the meter to "INPUT". Do not press the 'MENU'
button after each change within the submenu or the meter will advance to
the next menu item.
Every underlined "0" or "1" state is the factory preset value.
9-6
MIN/MAX*
BUTTON
SUB MENU
CONDITION
LOCKOUT CONFIGURATION #4
"L4C.1=0"
BAUD (communication rate) can be chosen.
"L4C.1=1"
BAUD (communication rate) cannot be
accessed.
"L4C.2=0"
SER.CNF (parity/stop-bit length) is
selectable.
"L4C.2=1"
SER.CNF (parity/stop-bit length) is not
selectable.
"L4C.3=0"
ADDRES (meter address # on a multipoint
bus) can be changed.
"L4C.3=1"
ADDRES (meter address # on a multipoint
bus) cannot be accessed.
"L4C.4=0"
DAT FT & BUS FT (format of data stream
and bus interaction for digital
communications) can be altered.
"L4C.4=1"
DAT FT & BUS FT (format of data stream
and bus interaction for digital
communications) cannot be accessed.
"L4C.5=0"
SERCNT (interval # of readings for the
automatic digital output of meter) can be
changed.
"L4C.5=1"
SERCNT (interval # of readings for the
automatic digital output of meter) cannot be
accessed.
"L4C.6=0"
Analog output trim input can be entered.
"L4C.6=1"
Analog output trim input cannot be entered.