METHOD III: SCALING BY SIMULATING A KNOWN LOAD WITH THREE OR MORE
DECIMAL POINT POSITIONS.
EXAMPLE: An end user has a displacement sensor with 5 inches of
travel and a 500 ohm output. You want to display in one ten
thousandths of an inch (desired display 0.0000 to 5.0000). Proceed
to Section 8.4 performing A, B and C sections.
1. Prior to simulating a known load you must set up the unit as
shown below.
Pressing the 'MENU' button will step you through all the different
menus listed below. When you wish to enter a menu, to the
submenu 1 stage, press the 'MIN' button. To make a selection, in
submenu 1, press the 'MAX' button. If the submenu has more then
one item to select press the 'MIN' button to scroll to the next item
of choice and again press the 'MAX' button to make your selection.
After making all your choices, in submenu 1, enter these selections
by pressing the 'MENU' button which will momentarily flash
"STORED", on the display.
If the item selected in submenu 1 has a submenu 2 associated with
it, press the 'MIN' button to enter the submenu 2. Select your
choice by pressing the 'MAX' button and enter your choice by
pressing the 'MENU' button.
Pressing the 'RESET' button one time brings you back one menu
item, pressing the 'RESET' button two times returns you to the run
mode.
WARNING! this also results in a hard reset of your peak and valley
and setpoints.
Proceed through the flow chart below entering the values listed in
submenu 1 and 2 columns.
8-49
8
Scaling to
Display
Engineering
Units