9.2.14 SP DB (Setpoint Deadband)

Refer to Section 10 for an in-depth discussion of these features.
Setpoint deadband is used to select the amount of hysteresis for the
setpoints (programmable from "0000" to "9999").
MENU
BUTTON
MAIN MENU
"SP DB":
* Press the 'MIN' button to show the value on the display.
The 'MIN' button also allows you to sequence through the digits of the
number being changed.
The 'MAX' button changes the value of the digit to be displayed.
The 'MENU' button stores the selected values for each "SP DB" condition
changed and advances the meter to the next configuration ("AL DB"):
The underlined item is the factory preset value.

9.2.15 AL DB (Alarm Deadband)

Refer to Section 10 for an in-depth discussion of these features.
Alarm deadband is used to select the amount of hysteresis for the alarms
(programmable from "0000" to "9999").
MENU
BUTTON
MAIN MENU
"AL DB":
* Press the 'MIN' button to show the value on the display.
The 'MIN' button also allows you to sequence through the digits of the
number being changed.
The 'MAX' button changes the value of the digit to be displayed.
Press the 'MENU' button to store the changes and advances the meter to
the next configuration ("OUT.CNF").
The underlined item is the factory preset value.
Meter Function Menus
MIN/MAX/MENU *
BUTTON
SUB MENU
"0000"
("0020")
MIN/MAX/MENU *
BUTTON
SUB MENU
"0000"
("0020")
DESCRIPTION
SETPOINTS 1 & 2 DEADBAND
Hysteresis for (with system decimal
points) Setpoints1 and 2
(Programmable from"0000" to "9999")
DESCRIPTION
ALARMS 1 & 2 DEADBAND
Hysteresis for (with system decimal
points) Alarms 1 & 2 (Programmable
from "0000" to "9999")